Portable Combustion Analyzer
Operation and Maintenance Manual
Instruction 0024-9472
Rev. 2 – October 2011
Product Leadership • Training • Service • Reliability
PCA 3 Manual
WARRANTY POLICY
Bacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from
defects in material and manufacture and will conform substantially to Bacharach Inc.’s
applicable specifications. Bacharach’s liability and Buyer’s remedy under this warranty are
limited to the repair or replacement, at Bacharach’s option, of this Product or parts thereof
returned to Seller at the factory of manufacture and shown to Bacharach Inc.’s reasonable
satisfaction to have been defective; provided that written notice of the defect shall have
been given by Buyer to Bacharach Inc. within two (2) years after the date of delivery of this
Product by Bacharach, Inc. The Oxygen (O ) sensor has a limited warranty of one (1) year.
2
Routine service and calibration are excluded from this warranty.
Bacharach, Inc. warrants to Buyer that it will convey good title to this Product. Bacharach’s
liability and Buyer’s remedy under this warranty of title are limited to the removal of any title
defects or, at the election of Bacharach, to the replacement of this Product or parts thereof
that are defective in title.
THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN AND ACCEPTED
IN LIEU OF (I) ANY AND ALL OTHER WARRANTIES, EXPRESS OR IMPLIED,
INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE: AND (II) ANY
OBLIGATION, LIABILITY, RIGHT, CLAIM OR REMEDY IN CONTRACT OR TORT,
WHETHER OR NOT ARISING FROM BACHARACH’S NEGLIGENCE, ACTUAL OR
IMPLIED. The remedies of the Buyer shall be limited to those provided herein to the
exclusion of any and all other remedies including, without limitation incidental or
consequential damages. No agreement varying or extending the foregoing warranties,
remedies or this limitation will be binding upon Bacharach, Inc. unless in writing, signed by
a duly authorized officer of Bacharach.
Register Your Warranty by Visiting
www.MyBacharach.com
NOTICE
Product improvements and enhancements are continuous, therefore the specifications and
information contained in this document may change without notice.
Bacharach, Inc. shall not be liable for errors contained herein or for incidental or
consequential damages in connection with the furnishing, performance, or use of this
material.
No part of this document may be photocopied, reproduced, or translated to another
language without the prior written consent of Bacharach, Inc.
Copyright © 2011, Bacharach, Inc., all rights reserved.
BACHARACH, PCA, B-SMART, and FYRITE are registered trademarks of Bacharach, Inc.
All other trademarks, trade names, service marks and logos referenced herein belong to
their respective companies.
2 0024-9472 Rev 2
PCA 3 Manual
TABLE OF CONTENTS
CHAPTER 1. INTRODUCTION ........................................................................ 7
1.1. About This Manual ....................................................................................... 7
1.2. Conventions ................................................................................................. 7
1.3. Safety ........................................................................................................... 7
1.4. General Description ..................................................................................... 9
1.5. Operation Overview ..................................................................................... 9
®
1.5.1. Turning On the PCA 3 ...................................................................... 9
1.5.2. The Combustion Test ..................................................................... 10
®
1.5.3. Turning Off the PCA 3 .................................................................... 10
1.6. Features and Benefits ................................................................................ 11
1.7. Sales Combo and Model Configurations .................................................... 12
1.8. Specifications ............................................................................................. 13
CHAPTER 2. HARDWARE ............................................................................ 15
2.1. Overview .................................................................................................... 15
2.2. Front Panel Buttons ................................................................................... 16
2.3. Backlit Display ............................................................................................ 17
2.4. Battery Compartment ................................................................................ 17
2.5. AC Adapter Connector (POWER) ................................................................ 17
2.6. Probe Connectors (Gas, Pressure, T-Stack) ................................................ 18
2.7. Differential Pressure Connector (∆P) ......................................................... 19
2.8. Primary Air Thermocouple Connector (T-AIR) ........................................... 19
2.9. Computer Interface (USB) .......................................................................... 19
2.10. Wireless Printer Port (IrDA) ..................................................................... 19
2.11. Options Connector (OPT) ......................................................................... 19
CHAPTER 3. INITIAL SETUP ......................................................................... 20
3.1. Overview .................................................................................................... 20
3.2. Power ......................................................................................................... 20
3.3. Connecting the Probe and Hose Assembly ................................................ 21
®
3.4. Preparing to Configure the PCA 3 .............................................................. 21
CHAPTER 4. CONFIGURATION PROCEDURES .............................................. 23
4.1. Default Operating Parameters ................................................................... 23
4.2. Fuel Selection ............................................................................................. 23
4.3. Auto/Manual CO Zero Selections ............................................................... 24
4.4. Temperature Units Selection ..................................................................... 25
4.5. Pressure Units Selection ............................................................................ 26
4.6. Pollution Units Selection ............................................................................ 27
0024-9472 Rev 2 3
PCA 3 Manual
4.7. Date Setup .................................................................................................. 27
4.8. Time Setup ................................................................................................. 28
4.9. O Reference Setup .................................................................................... 29
2
4.10. Print Pressure Selection ........................................................................... 30
4.11. Zoom Display Selection ............................................................................ 30
4.12. Logging Selection ..................................................................................... 31
4.13. Button Sound Selection ............................................................................ 31
4.14. Test ID Information .................................................................................. 32
4.14.1. Entering or Editing a Test ID ........................................................ 32
4.14.2. Selecting a Test ID ....................................................................... 33
4.14.3. Clearing Test ID Information........................................................ 33
4.15. User Name ............................................................................................... 34
4.15.1. Entering a User Name .................................................................. 34
4.15.2. Clearing a User Name .................................................................. 35
4.16. Language .................................................................................................. 36
4.17. Cal Reminder Period ................................................................................ 36
4.18. Run/Hold Screen Format .......................................................................... 37
CHAPTER 5. OPERATION ............................................................................ 40
5.1. Operating Tips ............................................................................................ 40
5.2. Turning On the Analyzer ............................................................................ 41
5.3. Low Battery Alarm ..................................................................................... 42
5.4. Locating Sampling Points ........................................................................... 42
5.5. Performing a Combustion Test .................................................................. 43
5.6. Selecting a Pressure Label .......................................................................... 45
5.7. Temperature Label Selection and Measurement ...................................... 46
5.8. Performing a Draft / Pressure Measurement ............................................ 47
5.9. Saving Test Data ......................................................................................... 48
5.10. Ending a Combustion Test........................................................................ 49
5.11. Emptying the Water Trap ......................................................................... 49
5.12. Data Logging ............................................................................................. 50
5.12.1. Overview ...................................................................................... 50
5.12.2. Turning On Data Logging ............................................................. 51
5.12.3. Setting the Logging Interval and Duration ................................... 52
5.12.4. Starting the Data Logging Process ............................................... 53
5.12.5. Ending the Data Logging Process ................................................. 53
5.13. Memory Functions ................................................................................... 54
5.13.1. Overview ...................................................................................... 54
5.13.2. Recalling Combustion Test Data .................................................. 54
5.13.3. Recalling Logged Test Data .......................................................... 55
5.13.4. Clearing Memory ......................................................................... 56
4 0024-9472 Rev 2
PCA 3 Manual
5.14. Downloading Stored Data to a Computer ................................................ 57
5.15. Importing Saved Data into a Spreadsheet ............................................... 57
5.16. Printing Test Data ..................................................................................... 59
5.17. Turning Off the Analyzer and Purging ...................................................... 61
CHAPTER 6. CALIBRATION .......................................................................... 62
6.1. B-Smart Sensors ......................................................................................... 62
6.2. Starting a Calibration ................................................................................. 62
®
6.3. B-Smart Sensor Replacement and Calibration .......................................... 63
6.4. Pressure Sensor Calibration ....................................................................... 64
6.4.1. Material Required .......................................................................... 64
6.4.2. Procedure ...................................................................................... 64
6.5. T-Stack Calibration ..................................................................................... 65
6.5.1. Material Required .......................................................................... 66
6.5.2. TS-Zero Procedure ......................................................................... 66
6.5.3. TS-Span Procedure ........................................................................ 67
6.6. T-Air Calibration ......................................................................................... 67
6.6.1. Material Required .......................................................................... 67
6.6.2. TA-Zero Procedure......................................................................... 67
6.6.3. TA-Span Procedure ........................................................................ 69
6.7. CO Sensor Calibration ........................................................................... 69
LOW
6.7.1. Material Required .......................................................................... 69
6.7.2. Procedure ...................................................................................... 70
6.8. SO Sensor Calibration ............................................................................... 72
2
6.8.1. Material Required .......................................................................... 72
6.8.2. Procedure ...................................................................................... 72
6.9. NO Sensor Calibration ................................................................................ 73
6.9.1. Material Required .......................................................................... 73
6.9.2. Procedure ...................................................................................... 73
6.10. NO Sensor Calibration ............................................................................. 74
2
6.10.1. Material Required ........................................................................ 74
6.10.2. Procedure .................................................................................... 74
6.11. CO Sensor Calibration ......................................................................... 76
HIGH
6.11.1. Material Required ........................................................................ 76
6.11.2. Procedure .................................................................................... 76
CHAPTER 7. MAINTENANCE ....................................................................... 78
7.1. Overview .................................................................................................... 78
®
7.2. PCA 3 Disassembly ..................................................................................... 78
7.2.1. Tools Required ............................................................................... 78
7.2.2. Procedure ...................................................................................... 78
0024-9472 Rev 2 5
PCA 3 Manual
7.3. Emptying the Water Trap Chamber ........................................................... 83
7.4. Replacing the Filter Element ...................................................................... 83
7.4.1. Material Required .......................................................................... 83
7.4.2. Procedure ...................................................................................... 83
7.5. Sensor Replacement .................................................................................. 84
7.6. Nitric Oxide (NO) Sensor Battery Replacement ......................................... 86
7.6.1. Material Required .......................................................................... 86
7.6.2. Procedure ...................................................................................... 86
7.7. Cleaning the Probe ..................................................................................... 87
7.7.1. Equipment Required ...................................................................... 87
7.7.2. Procedure ...................................................................................... 87
CHAPTER 8. TROUBLESHOOTING ............................................................... 88
8.1. Error Symbols ............................................................................................. 88
8.2. Diagnostics and Status Screens .................................................................. 88
8.3. Accessing the Diagnostic Menu ................................................................. 89
8.4. Accessing the Status Menu ........................................................................ 89
8.5. Error Messages (After Warm-up) ............................................................... 89
8.6. Replacement Parts ..................................................................................... 90
8.7. Accessories ................................................................................................. 93
8.8. Analyzer Repair .......................................................................................... 95
8.9. Service Centers ........................................................................................... 95
CHAPTER 9. DECLARATION OF CONFORMITY ............................................. 96
6 0024-9472 Rev 2
PCA 3 Manual
CHAPTER 1. INTRODUCTION
1.1. About This Manual
®
Thank you for investing in a Bacharach PCA 3 Combustion Analyzer. To assure
®
operator safety and the proper use of the PCA 3, please read the contents of this
manual for important information on the operation and maintenance of the
analyzer.
1.2. Conventions
WARNING: A warning statement denotes a potential hazard
associated with the use of this equipment. Failure to follow this
information could result in serious personal injury or death.
CAUTION: A caution statement indicates a potentially hazardous
situation which, if not avoided, may result in minor or moderate
injury. Caution statements may also be used to alert against unsafe
practices.
IMPORTANT: An important statement provides emphasis of an
important feature, operation, etc. Failure to follow this information
could void your warranty, result in improper operation, or cause
equipment damage.
NOTE: A note statement provides emphasis of a feature, operation,
practice, etc.
1.3. Safety
WARNING: This analyzer is not intended to be used as a safety
device.
WARNING: When testing an appliance, a full visual inspection of
the appliance should be performed to ensure its safe operation.
CAUTION: This analyzer is not intended to be used on a
continuous basis.
0024-9472 Rev 2 7
PCA 3 Manual
CAUTION: Do not store instrument or its sensors with solvents or
products that contain solvents.
CAUTION: Except for sensor and battery replacement, this
analyzer should only be opened and/or serviced by authorized
Bacharach personnel. Failure to comply may void the warranty.
HAZARDOUS AREA WARNING: This instrument has not been
designed to be intrinsically safe for use in areas classified as
hazardous locations. For your safety, DO NOT use it in hazardous
(classified) locations.
CAUTION: When the instrument is used in an inefficient furnace or
boiler application where there is a high emission of soot, the probe’s
sample filter may become clogged. Before every use check the filter
to confirm that it is clean or replace it with a new filter.
To prevent this from occurring, a smoke test should be performed
before operating under such conditions. This ensures that the
furnace or boiler is burning at a level appropriate for the use of this
instrument.
When the CO level exceeds the allowable threshold, a warning will
2
appear prompting the user to consider performing a smoke test (see
Figure 1-1). This screen is removed by pressing the ENT button.
Once the warning is cleared, it will not be displayed again for that
particular test. If a new test is started (by pressing the HOLD
button), the warning will be displayed again if the limit has been
exceeded.
Figure 1-1. High CO2 Warning
8 0024-9472 Rev 2
PCA 3 Manual
1.4. General Description
®
The PCA 3 is a commercial-grade hand-held combustion and emissions
analyzer designed for on-demand sampling of light industrial, institutional,
commercial and residential furnaces, appliances, and boilers.
The basic instrument is supplied with
a probe and hose assembly,
instruction manual, factory calibrated
smart sensors, 4 ‘AA’ alkaline
®
batteries, boot, Fyrite User
Software, USB cable, and carrying
case. Refer to Figure 1-2.
®
The PCA 3 can measure up to four
gases simultaneously, so it is the
perfect tool for service technicians,
inspectors, and boiler contractors
who need to determine combustion
efficiency, excess air, stack gas O
2
and CO levels, stack temperature,
®
draft, and differential pressure.
Figure 1-2. A Typical PCA 3 Kit
The analyzer can also directly measure and display NO, NO and SO with the
2 2
installation of the appropriate sensors. Combustion efficiency calculations can be
conducted for a variety of fuels (see Specifications later in this chapter).
®
B-Smart sensor technology allows a new sensor to be installed in the field
without having to calibrate the analyzer with gas before use.
Advanced data storage and communication features allow the operator to store
up to 500 individual combustion test records, which can later be recalled for
viewing, printing, or downloading to a personal computer. In its data logging
mode, the analyzer can store an additional 500 data logged records. The
optional AC power adapter allows the analyzer to run for extended periods of
time while data logging.
The optional sample conditioning probe is recommended when measuring NO
2
and SO to ensure the highest degree of measurement accuracy.
2
A probe and hose assembly with an integral thermocouple and filter/water trap
connect to the bottom of the analyzer, providing the means of drawing in gas
samples and for measuring stack temperature and draft.
1.5. Operation Overview
®
1.5.1. Turning On the PCA 3
®
The PCA 3 is turned ON by pressing its red I/O button. A warm-up period of 60
seconds then begins, during which time the analyzer performs self diagnostics.
At the end of the warm-up period, if no errors were detected the instrument will
display the Combustion Test HOLD screen. If errors were detected, the message
0024-9472 Rev 2 9
PCA 3 Manual
“ERRORS DETECTED” is displayed along with a list of the errors. These errors
must be corrected before proceeding with the combustion test.
Before starting a test be sure to select the fuel being burned. The default fuel
selected is Natural Gas. Note that the name of the fuel being burned is indicated
at the top of the run/hold screen.
1.5.2. The Combustion Test
To assure correct combustion efficiency calculations, the analyzer must know the
burner’s primary-air temperature. The analyzer normally uses its internal
temperature sensor for the primary-air temperature value, but this method is only
acceptable if the burner is using ambient room air. If the burner is drawing in cold
outside air, we recommend that the optional T-AIR thermocouple be used. This
thermocouple plugs into the bottom of the analyzer and is placed in the burner’s
primary-air stream.
Begin the combustion test by first inserting the analyzer’s probe tube into the
stack-gas stream of the appliance under test, and then pressing the RUN/HOLD
button to display the Combustion Test RUN screen. The analyzer will begin to
continuously monitor the stack temperature, %O and emission levels in the
2
stack gas and then display measured and calculated values on its display. The
recommended time required to achieve a stable measurement is a minimum of 3
minutes.
During a test, the CO sensor is protected from high CO levels by being
LOW
automatically flushed with fresh air when the detected CO level exceeds
4,000 ppm. The analyzer will automatically start using its optional CO sensor,
HIGH
if installed, at CO levels starting at 4,001 ppm, thus providing continuous CO
readings up to 20,000 ppm.
A keypad backlight enables a user to read the keypad in dimly-lit areas. Turn the
keypad backlight ON and OFF by briefly pressing the I/O button.
®
1.5.3. Turning Off the PCA 3
The analyzer is turned OFF by pressing and holding down the I/O button for at
least 2 seconds. Note that there is a 5-second delay before the analyzer actually
turns OFF, during which time the analyzer can be turned back ON by pressing
the RUN/HOLD button. In addition, there is a gas-purge feature that keeps the
analyzer’s pump running if the gas level inside the sensor chambers is
abnormally high at shutdown. With the probe removed from the stack and
sampling fresh air, the analyzer purges itself until the detected gas
concentrations drop below predetermined levels.
10 0024-9472 Rev 2
PCA 3 Manual
1.6. Features and Benefits
• Powered by 4 ‘AA’ alkaline batteries or NiMH rechargeable batteries. An
optional AC adapter provides extended operation.
• O and CO measurement standard. Optional measurement of up to two
2 LOW
additional gases, including CO , NO, NO , or SO .
HIGH 2 2
With the appropriate sensors installed, the analyzer optionally displays
•
pollution conversions for CO, NO, NO , and SO . Pollution conversions
2 2
3
include ppm, #/MBTU, mg/m , and g/GJ.
®
• B-Smart sensor technology allows pre-calibrated sensors to be installed in
the field. Sensors are provided with data that can be entered through the
®
PCA 3 software, or instrument calibration menus, for easy calibration.
• Automatic flushing of the CO sensor with fresh air if the CO level
LOW
exceeds 4,000 ppm, thus protecting the CO sensor from high CO levels.
LOW
To measure CO levels above 4,000 ppm, the analyzer automatically
switches to its CO sensor, if installed.
HIGH
Automatic purging of the gas-sample system if the detected gas levels are
•
abnormally high when the analyzer is turned OFF.
• Displays temperatures in either °F or °C
• Displays pressure in either inwc, mb, Pa, or hPa
• Backlit color graphic LCD with zoom capabilities
• Low battery alarm
Stores 500 individual combustion records, which can later be recalled for
•
viewing, printing, or downloading to a personal computer. Stores an
additional 500 data logged records.
Wireless IrDA link for printing current and stored combustion records,
•
pressure records, calibration data, and diagnostic data
USB connectivity for downloading data to personal computer
•
Field replaceable sensors and thermocouple
•
Two year warranty on analyzer and all gas sensors except the O sensor
• 2
which has a one (1) year warranty.
• Language options including English, French, and Spanish
• Custom Display Formats
®
Calibration Reminders: PCA 3 can be configured to remind the user that
•
calibration is past due.
• Auto/manual zeroing option allows the operator to select the start-up mode.
Auto-zero determines the zero reading of the CO channel in fresh air and
provides an offset for the CO measurements. Also, all sensing channels are
automatically zeroed on ambient air when the analyzer is first turned ON.
Manual zero is used to detect CO that may be present during start-up.
0024-9472 Rev 2 11
PCA 3 Manual
1.7. Sales Combo and Model Configurations
Sales Combo
(0024-xxxx)
Sales Combo Kit
(0024-xxxx)
Model Type
®
PCA 3 Only Part Number
(0024-xxxx)
Measurements
Oxygen (O ) • • • • • • •
2
Stack Temperature • • • • • • •
Primary/Ambient Air Temperature • • • • • • •
Carbon Monoxide Low (CO ) • • • • • • •
LOW
Pressure/Draft • • • • • • •
Carbon Monoxide High (CO ) • •
HIGH
Nitric Oxide (NO) • • • •
Nitrogen Dioxide (NO ) •
2
Sulfur Dioxide (SO ) • •
2
Calculations
Combustion Efficiency • • • • • • •
Excess Air • • • • • • •
Carbon Dioxide (CO ) • • • • • • •
2
NO (NO = NO + NO ) •
x x 2
NO referenced to % O •
x 2
CO referenced to % O • • • • • • •
2
NO referenced to % O • • • •
2
NO referenced to % O •
2 2
SO referenced to % O • •
2 2
12 0024-9472 Rev 2
7320 225 8447 8440
7321 235 8448 8441
7322 245 8449 8442
7323 255 8450 8443
265 8451 8444
7324
7325 275 8452 8445
7326 285 8453 8446
PCA 3 Manual
1.8. Specifications
Measurement (Based on Installed Sensors) Display Range
Oxygen 0.1 to 20.9%
Stack Temperature -4 to 2,192 °F (-20 to 1,200 °C)
Primary/Ambient Air Temperature -4 to 999 °F (-20 to 537 °C)
Carbon Monoxide (CO) (H compensated) 0 to 4,000 ppm
2
Pressure/Draft ±72” H O (±179 mb)
2
CO High Range 4,001 to 20,000 ppm
Nitric Oxide (NO) 0 to 3,000 ppm
Nitrogen Dioxide (NO ) 0 to 500 ppm
2
Sulfur Dioxide (SO ) 0 to 5,000 ppm
2
Calculated Value (Based on Installed Sensors) Display Range
Combustion Efficiency 0.1 to 100%
Excess Air 1 to 250%
Carbon Dioxide (dry basis) 0.1 to fuel-dependent max in %
NO (NO = NO + NO ) 0 to 3,500 ppm
x x 2
NO referenced to %O 0 to 9,999 ppm
x 2
CO referenced to %O 0 to 9,999 ppm
2
NO referenced to %O 0 to 9,999 ppm
2
NO referenced to %O 0 to 9,999 ppm
2 2
SO referenced to %O 0 to 9,999 ppm
2 2
NOTE: Calculations are performed only when the measured
oxygen level is below 16.0% and the stack temperature is below
2,000 °F (1,093 °C).
0024-9472 Rev 2 13
PCA 3 Manual
Reading Performance Accuracy
±0.3% O on practical concentrations of stack gas
2
O
2
(mix of O , CO , and N )
2 2 2
Greater of ±5% of reading or ±10 ppm, between 0-2,000 ppm,
CO
±10% of reading between 2,001 and 20,000 ppm.
NO Greater of ±5% of reading or 5 ppm
NO Greater of ±5% of reading or ±5 ppm between 0-500 ppm
2
SO Greater of ±5% of reading or ±10 ppm between 0-2,000 ppm
2
±4 °F (±2 °C) between 32 and 255 °F (0 and 124 °C)
Stack Gas Temp ±6 °F between 256 and 480 °F (±3 °C between 125 and 249 °C)
±8 °F between 481 and 752 °F (±4 °C between 250 and 400 °C)
Primary/Ambient
±2 °F between 32 and 212 °F (±1 °C between 0 and 100 °C)
Air Temp
±0.02 inches from -1 to 1 inwc;
Pressure/Draft ±2% of reading from -10 to 10 inwc;
±3% of reading from -40 to 40 inwc
System Flow Rate
200 cc/min minimum
with Probe
General Specification Description
Dimensions (H x W x D) 9.0 x 3.0 x 2.5 inches (22.9 x 7.6 x 6.3 cm)
Analyzer with Batteries: ........................... 1.4 lb (0.6 kg)
Weight
Probe/Hose Assembly: ............................ 1.0 lb (0.5 kg)
Natural Gas Wood
• •
Coal Kerosene
• •
Fuels Available for
Oil #2 Bagasse
• •
Combustion Calculations
Oil #4 Propane
• •
Oil #6 Digester Gas
• •
Warm-up Time 60 seconds (Sensors checked and auto zeroed)
Alkaline: .............................. 10 hours of operation (min)
4 “AA” Batteries
(Disposable)
NiMH Rechargeable: ................... Operating time varies
Power
AC Adapter
100-240 VAC; 50/60 Hz: ....................... Continuous use
(Optional)
Analyzer: ................................ 32 to 104 °F (0 to 40 °C)
Operating Temperature
Probe Tip: ..................................1,472 °F (800 °C) Max.
Operating Humidity Analyzer: ...................... 15 to 90% RH, non-condensing
Analyzer: ................................................... Atmospheric
Operating Air Pressure
Probe: ............... 10″ H O (25 mb) draft max. at probe tip
2
500 complete combustion test records
Memory
500 complete logged combustion test records
Printer: .......................... Infrared (IrDA) communications
Interfaces
Computer: ........................... USB 2.0 (mini-B connector)
14 0024-9472 Rev 2
PCA 3 Manual
CHAPTER 2. HARDWARE
2.1. Overview
Figure 2-1. Front and Bottom Views Showing Key Components
0024-9472 Rev 2 15
PCA 3 Manual
2.2. Front Panel Buttons
®
The PCA 3 is controlled by various buttons, while a color graphic LCD (with
automatic dimming feature) is used to display all combustion and emission test
data and analyzer parameters. Descriptions of the front panel buttons are given
below. Note that a button may perform multiple functions depending on the
current screen that is displayed.
PRINT Transmits the data displayed on the screen to a printer
through the IrDA communications port.
F1
PAGE- When viewing the Memory or Logging Directory, each
press of this button pages down through the directory.
Holding this button down speeds up the paging process.
MENU Displays the Main Menu.
ZERO When viewing the Pressure screen, this button zeros the
F2
pressure sensor to current atmospheric conditions. When
viewing the Temperature screen, this button zeroes the
temperature channel difference.
SAVE Saves the data currently displayed on the LCD in
memory. Up to 500 individual Combustion Test,
Temperature, and Pressure records can be saved. After
500 records have been saved, the memory must be
F3 cleared to continue saving additional data. The analyzer
will not overwrite old data.
PAGE+ When viewing the Memory or Logging Directory, each
press of this button pages through the directory. Holding
this button speeds the paging process.
The arrow buttons move the cursor on the LCD. In screens that
▲
require the entry of alphanumerical data, use the ◄► buttons to
▼
move cursor across the screen and then use the ▲▼ buttons to
◄
increment and decrement the data. When viewing a menu, use the
►
◄► buttons to quickly move to the top and bottom of the menu.
Selects a highlighted item. In addition, if changes were made to
one of the analyzer’s operating parameters (e.g., date, time, O
2
ENT
reference, etc.), pressing this button confirms those changes and
saves them in memory.
Starts and stops a combustion test when the Combustion Test
screen is displayed. Pressing this button in any other screen
RUN/ returns the analyzer to the Combustion Test HOLD screen.
HOLD
Pressing this button during the 5 second turn-off-delay period will
abort the turn-off process and also return the analyzer to the
Combustion Test HOLD screen.
Displays a previously viewed screen. In addition, if changes were
made to one of the analyzer’s operating parameters (e.g., date,
ESC
time, O reference, etc.), pressing this button aborts those
2
changes, restores the old values, and then displays the previously
viewed screen.
Turns the analyzer ON and OFF, and is also used to turn the
I/O
keypad LEDs ON and OFF.
16 0024-9472 Rev 2
PCA 3 Manual
For example, the functions of the F1, F2, and F3 buttons are defined by labels
appearing above them on the LCD. The labels that appear depend on the
functions that can be performed in the particular screen being displayed.
NOTE: A keypad backlight enables a user to read the keypad in
dimly-lit areas. Turn the keypad backlight ON and OFF by briefly
pressing the I/O button.
NOTE: When the analyzer is turned OFF, there is a 5-second
delay, during which time an operator can keep the analyzer turned
ON by pressing the RUN / HOLD button. Also note that if the
measured emission levels are above predetermined limits at the
time the instrument is turned OFF, the pump is automatically started
and purges the sensor compartment with fresh air until the gas
levels inside the analyzer are reduced. If desired, the purging
process can be aborted by again pressing the I/O button, though it is
not recommended.
2.3. Backlit Display
A large backlit color graphic display shows multiple combustion test values
simultaneously, and includes a zoom capability that provides extra large text.
Figure 2-2. Sample Display
2.4. Battery Compartment
®
The PCA 3 is powered by either its four internal batteries or by an optional AC
power adapter that operates from any convenient source of 100–240 VAC, 50/60
Hz power. Batteries may be either disposable alkaline or rechargeable NiMH.
2.5. AC Adapter Connector (POWER)
The AC power adapter P/N 0024-1254 can be used as an external power supply,
which will run the analyzer on a continuous basis. It uses 100-240 VAC at
50/60 Hz.
0024-9472 Rev 2 17
PCA 3 Manual
Figure 2-3. Battery Compartment
Figure 2-4. Bottom View Showing Connectors
2.6. Probe Connectors (Gas, Pressure, T-Stack)
The probe and hose assembly are connected to the analyzer by way of the
following connectors.
• Stack-gas thermocouple to the analyzer’s T-STACK connector
• Stack-gas hose to the analyzer’s GAS connector
• Draft hose to the analyzer’s +∆P connector
Observe that the probe connectors are of different sizes and shapes, which
prevents incorrect connection to their associated connectors on the analyzer.
See Figure 2-5.
18 0024-9472 Rev 2
PCA 3 Manual
Figure 2-5. Probe Connectors GAS, +∆P, and T-STACK
2.7. Differential Pressure Connector (∆P)
Draft is measured by connecting the probe’s draft hose to the +∆P fitting, while
leaving the -∆P fitting open to the atmosphere. See Figure 2-5.
In addition to measuring draft, the “+” and “-” ∆P fittings can also be used to
measure the differential pressure between two areas by first connecting a hose
(P/N 0024-1103) to the -∆P fitting, and then inserting the open end of this hose
into the area being used as the reference pressure. The analyzer’s probe is then
inserted into the area where differential pressure is to be measured.
2.8. Primary Air Thermocouple Connector (T-AIR)
If thermocouple P/N 0104-1797 (10 feet long) or Utility Wand P/N 0104-1799
(12-inch ridged probe with handle and 5 foot coiled cable) is to be used to
measure the burner’s primary air temperature, then connect either of these
thermocouples to the analyzer’s T-AIR connector. See Figure 3-2 on page 22.
2.9. Computer Interface (USB)
Data stored in the analyzer’s memory can be downloaded to a personal
computer by connecting USB data cable P/N 0104-4032 between the USB ports
®
of the computer and analyzer. The PCA 3’s USB port is on its right side. See
Figure 2-1 on page 15.
2.10. Wireless Printer Port (IrDA)
Data that has been stored in the analyzer’s memory can be printed on a
compatible IrDA (Infrared data associated) wireless printer by aligning their IrDA
®
communication ports. The IrDA port of the PCA 3 is located at the top of the
device. See Figure 2-1 on page 15.
2.11. Options Connector (OPT)
The options connector is used for optional external measurement features. Refer
to Figure 3-2 on page 22.
0024-9472 Rev 2 19
PCA 3 Manual
CHAPTER 3. INITIAL SETUP
3.1. Overview
®
Before configuring or using the PCA 3, you MUST:
• provide power
• connect the probe and hose assembly.
3.2. Power
®
Three options are available to power the PCA 3:
• disposable alkaline batteries
• rechargeable NiMH batteries
AC power adapter.
•
If you are using either alkaline or NiMH rechargeable batteries to power the
analyzer, follow the instructions below to install or replace the batteries.
Remove the battery cover from back of unit (Figure 3-1).
•
• Remove (and properly dispose of) any old batteries.
Install a set of four ‘AA’ alkaline or NiMH batteries, per the “+” and “-”
•
markings inside the battery compartment.
Replace the battery cover.
•
Figure 3-1. Battery Cover and Compartment
The AC power adapter is capable of powering the analyzer on a continuous
basis. The adapter plugs into an appropriate 100-240 VAC, 50/60 Hz wall outlet,
and produces an output of +6 VDC. The adapter’s output connector plugs into
the analyzer’s POWER jack located on the bottom of the unit (refer to Figure 3-2
on page 22).
20 0024-9472 Rev 2
PCA 3 Manual
3.3. Connecting the Probe and Hose Assembly
To attach the probe and hose assembly to the analyzer follow the steps below
and refer to Figure 3-2 on page 22.
1. Push the gas sample hose connector, the larger of the two connectors
(giving a slight twist), onto the analyzer’s GAS fitting.
2. Push the draft hose connector, the smaller connector (giving a slight
twist), onto the analyzers +∆P fitting.
3. Push the stack gas thermocouple connector into the T-STACK jack
(connector fits in only one way).
4. Push the optional primary/ambient air thermocouple into the T-AIR jack
(connector fits in only one way).
IMPORTANT: To assure the accurate calculation of
combustion efficiency, the optional primary/ambient air
thermocouple must be used when the burner’s primary-air
temperature is not the same as the room temperature.
5. Inspect all hoses for cracks. If any hose is found to be defective, replace
the entire probe and hose assembly. Check that the water trap is empty,
and the filter is not dirty or saturated with water.
®
3.4. Preparing to Configure the PCA 3
®
After initial setup is complete, turn on the PCA 3. For detailed power-up
®
instructions, refer to Turning On the PCA 3 on pages 9 and 41. You are now
®
ready to configure your PCA 3 for your particular application and to your
particular preferences. Refer to Configuration Procedures on page 23. You
®
configure the PCA 3 by navigating the menus and changing key parameter
values. It may be helpful to review Front Panel Buttons on page 16 in
®
preparation for navigating the menu interface of the PCA 3.
0024-9472 Rev 2 21
PCA 3 Manual
Figure 3-2. Connecting the Probe to the Analyzer
22 0024-9472 Rev 2
PCA 3 Manual
CHAPTER 4. CONFIGURATION PROCEDURES
4.1. Default Operating Parameters
®
The PCA 3 is set up at the factory for the following operating parameters. To
change any of these parameters, perform the associated procedure that follows.
Parameter Factory Default To Change Value, See…
Fuel Natural Gas Page 23
Auto/Manual CO Zero Auto Zero Page 24
Temperature Units °F Page 25
Pressure Units Inches of Water Column (inwc) Page 26
Pollution Units ppm Page 27
Date Current MM/DD/YY Page 27
Time Current EST HH:MM AM/PM Page 28
O Reference 0% Page 29
2
Print Pressure No Page 30
Zoom Standard Page 30
Logging No Page 31
Button Sound On Page 31
Test ID Information Page 32
User Name Page 34
Language English Page 36
Calibration Reminder Never Page 36
Run/Hold Screen
See page 37 for default Page 37
Format
4.2. Fuel Selection
To assure the accurate calculation of combustion efficiency, select the fuel being
burned as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight FUEL and then press ENT to display
the FUEL MENU.
3. Use the ▲▼ buttons to scroll through the list of available fuels until the
desired fuel is highlighted. In the example shown, PROPANE has been
selected (If custom fuels are added, they will be displayed at the bottom
of the list).
NOTE: Use the ◄► buttons to quickly scroll to the bottom
and top of the list.
0024-9472 Rev 2 23
PCA 3 Manual
4. Press ENT to save the selection and display the Combustion Test
HOLD screen. Observe that the name of the selected fuel should now
appear at the top of the screen.
NOTE: In addition to the standard fuels programmed into the
®
PCA 3, Bacharach can develop custom fuel codes based on the
®
customers specific needs. The PCA 3 can be programmed with 2
additional fuels which can be added to the instrument using the
Fyrite User Software (FUS). Consult factory for price and delivery.
Figure 4-1. Fuel Selection
4.3. Auto/Manual CO Zero Selections
®
Select how the PCA 3 performs a zero function.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight CO Zero Setting, and then press ENT
to display the CO ZERO SETTING MENU.
4. Use the ▲▼ buttons to highlight the desired zeroing option (Manual or
Auto-Zero). In the example shown, Manual Zero has been selected.
5. Place the instrument in fresh air to zero, and press ENT to start a 60-
second count down.
6. Press ENT to save and re-display the SETUP MENU.
Figure 4-2. Auto/Manual Zero Selection
24 0024-9472 Rev 2
PCA 3 Manual
®
NOTE: The PCA 3 performs a zero function during warm-up.
Among other purposes, an auto-zero determines the “zero reading”
of the CO channel in fresh air and provides an offset for the CO
measurement. A manual zero detects background CO during
startup.
Manual Zero: When the CO channel is set to manual zero, the analyzer does not
zero the CO sensor to ambient conditions during start up. In this
mode, the “fresh air zero” established during manual mode setup
is stored in memory and used for the calculation of CO.
Auto Zero: When the CO channel is set to auto zero, the CO sensor is zeroed
to the ambient CO level during start up. Important: When using
this mode, the analyzer must be turned ON in fresh air; otherwise,
incorrect CO readings will occur.
Auto Zero determines the zero reading of the CO channel in fresh air and
provides an offset for the CO measurements. Manual zero is used to detect CO
that may be present during start-up. The auto/manual zeroing option allows the
operator to select the start-up mode. It is selectable from the CO Zero Setting of
the Setup Menu.
NOTE: Both methods take the user through a 60-second count
down to establish a new zero. The manual mode establishes a
fresh air zero and stores it for use during instrument startup.
4.4. Temperature Units Selection
Select to display temperature in either °F or °C as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight TEMP UNITS, and then press ENT to
display the TEMP UNITS MENU.
4. Use the ▲▼ buttons to highlight the desired temperature units. In the
example shown, Fahrenheit has been selected.
5. Press ENT to save the selection and re-display the SETUP MENU.
0024-9472 Rev 2 25
PCA 3 Manual
Figure 4-3. Temperature Units Selection
4.5. Pressure Units Selection
Select to display pressure in Inches of Water Column (inwc), millibar (mb),
Pascals (Pa), or hectoPascals (hPa) as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight PRESSURE UNITS, and then press
ENT to display the PRESSURE UNITS MENU.
4. Use the ▲▼ buttons to high light the desired pressure units. In the
example shown, InchesWater has been selected.
5. Press ENT to save the selection and re-display the SETUP MENU.
Figure 4-4. Pressure Units Selection
26 0024-9472 Rev 2
PCA 3 Manual
4.6. Pollution Units Selection
®
The PCA 3 is capable of converting the measured ppm levels of CO, NO, NO ,
2
and SO to various pollution units using CFR40 Part 60 emission factors. Note
2
that the pollution unit conversions for NO, NO and NOx are based on the
2
molecular weight of NO .
2
®
Configure the PCA 3 to display pollution units in parts per million (ppm), pounds
of pollutant per million BTU (#/Mbtu), milligrams of pollutant per cubic meter of
3
gas (mg/m ), or grams of pollutant per gigajoule (g/GJ) as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight POLLUTION UNITS, and then press
ENT to display the POLLUTION UNITS MENU.
4. Use the ▲▼ buttons to highlight the desired pollution units. In th e
example shown, ppm has been selected.
5. Press ENT to save the selection and re-display the SETUP MENU.
Figure 4-5. Pollution Units Selection
4.7. Date Setup
The date is stored in the format: MM/DD/YY. Its value is part of the date and time
stamp that is saved along with each combustion test record. Set the analyzer’s
internal clock to the current date as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight DATE, and then press ENT to display
the DATE MENU.
4. First use the ◄► buttons to move the cursor across the screen until it is
over the digit to be changed, and then press the ▲▼ buttons until the
desired value is displayed.
0024-9472 Rev 2 27
PCA 3 Manual
5. Repeat Step 4 until the values for month, day, and year have been set.
6. Press ENT to save the selection and re-display the SETUP MENU, or
press ESC to abort this procedure and retain the old date values.
NOTE: The real time clock is powered by the main
batteries and is maintained by a coin cell battery on the
Main PCB in the absence of batteries. Bacharach
recommends changing the coin cell (P/N 0204-0020) every
5 years.
Figure 4-6. Date Setup
4.8. Time Setup
The time is stored in the format: hh:mm:ss AM/PM. Its value is part of the date
and time stamp that is saved along with each combustion test record. Set the
analyzer’s internal clock to the current time as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight TIME, and then press ENT to display
the TIME MENU.
4. First use the ◄► buttons to move the cursor across the screen until it is
over the digit to be changed, and then press the ▲▼ buttons until the
desired value is displayed.
5. Repeat Step 4 until values for hour, minute, and meridian are set.
NOTE: The value for seconds cannot be entered, but is
displayed and stored as part of the combustion test record.
6. Press ENT to save the displayed time values and re-display the SETUP
MENU, or press ESC to abort and retain the old time values.
28 0024-9472 Rev 2
PCA 3 Manual
Figure 4-7. Time Setup
4.9. O Reference Setup
2
The measured values of CO, NOx, and SO can be individually referenced to a
2
specific O percentage of between 0 and 15%. Individually set up the O
2 2
reference value for each of the above gases as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight O REF, and then press ENT to
2
display the O REFERENCE screen.
2
4. Use the ▲▼ buttons to highlight the desired measurement, and then
press ENT to display the REF TO O screen for that measurement. In
2
the example shown, CO has been selected.
5. First use the ◄► buttons to move the cursor across the screen until it is
over the digit to be changed, and then press the ▲▼ buttons until the
desired value is displayed.
6. Press ENT to save the displayed value and re-display the O
2
REFERENCE screen, or press ESC to abort this procedure and retain
the old O reference value.
2
7. If the O reference value for more than one gas is being set, repeat
2
Steps 4, 5, and 6 for each measurement.
Figure 4-8. O Reference Setup
2
0024-9472 Rev 2 29
PCA 3 Manual
4.10. Print Pressure Selection
Select whether to print or not print the pressure measurement on the combustion
test printout as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight PRINT PRESSURE, and then press
ENT to display the PRINT PRESSURE screen.
4. Use the ▲▼ buttons to highlight either No (do not print pressure) or Yes
(print pressure). In the example shown, Yes has been selected.
5. Press ENT to save the selection and re-display the SETUP MENU.
Figure 4-9. Print Pressure Selection
4.11. Zoom Display Selection
Combustion test data in the Run/Hold screen can be shown with enlarged
characters to make viewing easier. The operator can set zoom levels to
Standard, 2X, or 3X. The Standard zoom setting will display seven lines of
combustion test data at one time, 2X which will display five lines of combustion
test data with enlarged characters, and 3X which will display four lines of
combustion test data with enlarged characters. The operator can scroll through
the complete list of measured and calculated data no matter what zoom level has
been selected. Select desired zoom level as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight ZOOM, and then press ENT to display
the ZOOM screen.
4. Use the ▲▼ buttons to select the desired Zoom level. Options include
STANDARD, 2X, and 3X. Standard will display seven lines of
Combustion test data, 2X will display five, and 3X will display four.
5. Press ENT to save the selection and re-display the SETUP MENU.
30 0024-9472 Rev 2
PCA 3 Manual
Figure 4-10. Zoom Display Selection
4.12. Logging Selection
When the logging function is activated, up to 500 combustion test records will be
automatically stored in memory at a preset interval over a predetermined length
of time.
Refer to Data Logging starting on page 50 for detailed information on how to
select the logging function, how to set the interval and duration time periods, and
how to view or download the stored data.
Figure 4-11. Logging Selection
4.13. Button Sound Selection
The audible sound used to signal when a button is pressed can be turned OFF
and ON as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight BUTTON SOUND, and then press
ENT to display the BUTTON SOUND screen.
4. Use the ▲▼ buttons to highlight either OFF or ON. In the example
shown, OFF has been selected.
5. Press ENT to save the selection and re-display the SETUP MENU.
0024-9472 Rev 2 31
PCA 3 Manual
Figure 4-12. Button Sound Selection
4.14. Test ID Information
Test records can be identified (e.g., customer’s name, burner number, and
location) by manually entering up to three lines of text, with each line containing a
maximum of 20 alphanumeric characters. When a Test ID is selected this
information will be associated with all succeeding test records, and will appear at
the top of each test record when printed, and in CSV files when records are
downloaded to a PC. The chosen Test ID remains in effect until it is deselected,
a new Test ID is selected, or the instrument is turned off. Up to 30 Test IDs can
be pre-entered for later retrieval.
NOTE: To save time, Test ID information can be entered using the
Fyrite User Software.
4.14.1. Entering or Editing a Test ID
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight TEST ID, and then press ENT to
display the TEST ID menu.
4. Use the ▲▼ buttons to highlight EDIT TEST ID, and then press ENT to
display the EDIT TEST ID menu, which displays the first line of each
Test ID record.
5. Use the ▲▼ buttons to select which Test ID to edit and then press ENT
to display all three lines of that record. Periods (.....) are used to identify
empty lines.
6. Use the ▲▼ buttons to choose which of the three Test ID lines to edit
and then press ENT to begin editing the chosen line.
7. Use the ▲▼ buttons to enter the desired character and then move to
the next character position by pressing the right arrow key. Repeat this
step until the line is complete. A maximum of 20 characters can be
entered. Press ENT to accept the information.
32 0024-9472 Rev 2
PCA 3 Manual
8. At this time, either return to Step 6 to choose another line to edit, or end
this procedure by highlighting Edit Complete and pressing the ENT key
to return to the Edit Test ID Menu.
9. Press ESC to go to the Test ID Menu or the RUN/HOLD key to return to
the Run/Hold screen.
Figure 4-13. Editing a Test ID
4.14.2. Selecting a Test ID
1. From the Test ID menu, use the ▲▼ buttons to highlight SELECT
TEST ID, and then press ENT to display the SELECT TEST ID menu,
which displays the first line of each Test ID record.
Figure 4-14. Selecting a Test ID
2. Use the ▲▼ buttons to choose a pre-entered Test ID or choose NO
Test ID (located at bottom of list) to deselect a previously chosen
record, then press ENT to make the selection and return to the Test ID
menu.
3. Press ESC to end the procedure and return to the Setup Menu or press
the RUN/HOLD key to return to the Run/Hold Screen.
4.14.3. Clearing Test ID Information
1. From the Test ID menu, use the ▲▼ buttons to highlight CLEAR TEST
ID, and then press ENT to display the CLEAR TEST ID menu.
2. Do one of the following to clear Individual Records or All Records.
Individual Records: Use the ▲▼ buttons to highlight Individual
Records, then press ENT to display the Clear Individual menu. Again
use the ▲▼ buttons to highlight the individual record to clear, then
press ENT to clear the record.
0024-9472 Rev 2 33
PCA 3 Manual
All Records: Use the ▲▼ buttons to highlight All Records, then press
ENT to display the Clear All menu. Again use the ▲▼ buttons to
highlight Yes, then press ENT to clear all records.
3. Press ESC to end this procedure and return to the Test ID menu or
press the RUN/HOLD key to return to the Run/Hold screen.
Figure 4-15. Clearing a Test ID
4.15. User Name
The name of the user or owner of the analyzer (e.g., company name, address,
phone number) can be stored in memory by manually entering up to three lines
of text, with each line containing up to 20 alphanumeric characters. This
information will appear at the top of each printout until new information is entered
or cleared.
NOTE: To save time, user name information can be entered using
®
the Fyrite User Software.
4.15.1. Entering a User Name
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight USERNAME, and then press ENT to
display all three lines of the EDIT USERNAME screen. Periods(.....)
identify empty lines.
4. Use the ▲▼ buttons to choose which of the three Username lines to
edit, then press ENT to begin editing the chosen line.
5. Use the ▲▼ buttons to enter the desired character and then move to
the next character position by pressing the right arrow key. Repeat this
step until the line is complete. A maximum of 20 alphanumeric
characters can be entered.
6. Press ENT to accept the entered information.
34 0024-9472 Rev 2
PCA 3 Manual
7. At this time, either return to Step 4 to choose another line to edit, or end
this procedure by highlighting Edit Complete and pressing the ENT key
to return to he Setup Menu.
8. Press ESC to go to the Main Menu or the RUN/HOLD key to return to
the Run/Hold screen.
Figure 4-16. Entering a User Name
4.15.2. Clearing a User Name
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight USERNAME, and then press ENT to
display all three lines of the EDIT USERNAME screen. Periods (.....)
identify empty lines.
4. Use the ▲▼ buttons to choose which of the three Username lines to
clear, then press CLEAR (F3) to clear that line.
5. At this time, either return to Step 4 to choose another line to clear, or
end this procedure by highlighting EDIT COMPLETE and pressing the
ENT key to return tot he Setup Menu.
6. Press ESC to go to the Main Menu or the RUN/HOLD key to return to
the Run/Hold screen.
Figure 4-17. Clearing a User Name
0024-9472 Rev 2 35
PCA 3 Manual
4.16. Language
Information on the display screen can be shown in English, French, or Spanish.
Select the desired language as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight LANGUAGE SELECTION, and then
press ENT to display The Language Selection Menu.
4. Use the ▲▼ buttons to highlight the desired language, then press ENT
to activate the selection and re-display the Setup Menu.
Figure 4-18. Selecting a Language
4.17. Cal Reminder Period
The analyzer can be set to indicate a calibration reminder during the 60 second
warm-up period. Calibration reminders can be preset to occur never, 6, 8, 10,
12, or 15 months after the last calibration. When the preset period is exceeded
the instrument will display the reminder, and how long since the sensors were
last calibrated. The reminder will be displayed at the end of the 60 second warm-
up period. If a calibration reminder is displayed the operator can press the
RUN/HOLD key to move to the Run/Hold screen for normal operation. Regular
calibration periods of 6 months to a year for all gas sensors (except Oxygen) are
recommended. Set the calibration reminder period as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight CAL REMINDER PERIOD, and then
press ENT to display CAL Reminder Period Menu.
4. Use the ▲▼ buttons to highlight the desired CAL reminder period, and
then press the ENT key to activate the selection and re-display the
Setup Menu.
36 0024-9472 Rev 2
PCA 3 Manual
NOTE: The date and time settings must be correct to get accurate
calibration reminders.
Figure 4-19. Selecting the Calibration Reminder Period
4.18. Run/Hold Screen Format
®
PCA 3 test data is located in the Run/Hold screen. By pressing the
key, you should hear the pump running and see the word RUN at
RUN/HOLD
the upper left hand corner of the display. The instrument is continuously
measuring and calculating the data that is shown in the Run/Hold screen. Press
the RUN/HOLD key again, the pump should stop running and the word HOLD
should be shown at the upper-left hand corner of the display. The instrument will
now show the last measured and calculated data taken before the instrument
was placed in HOLD. Use the up and down arrow keys to scroll through the
complete list of measured and calculated values when the instrument is running
or in the hold mode. The default order in which data appears in the Run/Hold
screen is shown below.
O Oxygen
2
CO Carbon Monoxide
EFF Combustion Efficiency
CO Carbon Dioxide
2
T-STK Stack Temperature
T-AIR Ambient/Primary Air Temperature
EA Excess Air
CO(#) Carbon Monoxide content referenced to an Oxygen percentage
NO Nitric Oxide
NO Nitrogen Dioxide
2
NOx Oxides of Nitrogen (NO and NO combined)
2
SO Sulfur Dioxide
2
NO(#) Nitric Oxide content referenced to an Oxygen percentage
NO (#) Nitrogen Dioxide content referenced to an Oxygen percentage
2
NOx(#) Oxides of Nitrogen content referenced to an Oxygen percentage
SO (#) Sulfur Dioxide content referenced to an Oxygen percentage
2
0024-9472 Rev 2 37
PCA 3 Manual
NOTE: A number sign (#) denotes the current O Reference
2
selected.
NOTE: Stars (***) appear in measurement and calculation fields of
sensors that are not installed.
Change the order in which data is displayed as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight RUN/HOLD FORMAT, and then press
ENT to display Run/Hold Format Menu.
4. Use the ▲▼ buttons to highlight Edit Format, and then press the ENT
key to show the Edit Run/Hold Format, where the current order of
combustion data is displayed.
5. Change data shown for a particular location by first using the ▲▼
buttons to highlight the location. Then press the ENT key to select the
location; the cursor will start to flash.
6. Use the ▲▼ buttons to scroll through and select the desired data to
appear at that location. Press the ENT key to make the selection.
7. Change data displayed at other locations by repeating Steps 5 and 6.
8. When finished, use the ▼ button to highlight Edit complete at the
bottom of the list, then press ENT to save the new display format and
return to the Run/Hold Format screen.
Figure 4-20. Changing the Order in Which Data Is Displayed
Reset Display format back to factory default settings as follows.
38 0024-9472 Rev 2
PCA 3 Manual
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight RUN/HOLD FORMAT, and then press
ENT to display Run/Hold Format Menu.
4. Use the ▲▼ buttons to highlight RESET FORMAT, then press ENT to
display the Reset Format screen.
5. Use the ▲▼ buttons to select YES, then press ENT to reset the display
and return to the Run/Hold Format screen.
Figure 4-21. Resetting Display Formats Back to Factory Defaults
0024-9472 Rev 2 39
PCA 3 Manual
CHAPTER 5. OPERATION
5.1. Operating Tips
• When an analyzer is brought in from a cold vehicle, let it warm up slowly to
minimize condensation. Temperatures below freezing will not damage the
analyzer.
CAUTION: Although the analyzer itself is not damaged by an
extremely cold environment, the electrochemical sensors may
be damaged. The O sensor’s electrolyte will freeze at
2
approximately -20 ºF and the other sensors at approximately
-94 ºF. If the analyzer is exposed to an extremely cold
condition, it is strongly suggested that the sensor housings be
examined for hairline cracks. Be aware that a leaking sensor
can cause chemical burns to the skin and possibly damage the
PCB assemblies.
• Ensure that the analyzer is sampling fresh air when turned ON. Pulling a
stack gas sample through the analyzer during its warm-up period will not
damage the analyzer, but it will result in incorrect sensor readings, and may
result in sensor error messages appearing after the warm-up cycle
completes.
Note that flue gas condensate is acidic and very corrosive. It is important not
•
to allow the analyzer’s internal components to come in contact with
condensate for long periods of time.
• Before each use, inspect the filter element of the water trap / filter assembly.
Replace the filter if it looks dirty. See Replacing the Filter Element on page
83.
• When sampling flue gas, keep the analyzer above the water trap, and keep
the trap in a vertical position. This will maximize the effectiveness of the trap
and keep liquid condensate from being drawn directly into the analyzer.
When liquid condensate is seen inside the water trap, empty the trap before
•
it becomes full. See Emptying the Water Trap on pages 49 and 83.
• The analyzer should be purged after performing a combustion test. After
removing probe from the stack, let the pump run for at least 10 minutes to
remove any stack gases and dry any condensate from inside the sensor
chamber and probe assembly. If the analyzer is turned OFF with high levels
of flue gas remaining in the analyzer, then the pump will continue to run and
the message “PURGING SENSORS” will appear on the display until all flue
gas levels fall below predetermined levels.
• When storing the analyzer, it’s a good idea to empty the water trap and
leave it open to further dry it out.
Calibrate the analyzer every 6 months to 1 year to ensure its accuracy.
•
40 0024-9472 Rev 2
PCA 3 Manual
5.2. Turning On the Analyzer
Connect the probe and hose assembly, and make sure that the analyzer is
properly set up according to Initial Setup (page 20) and Configuration Procedures
(page 23).
IMPORTANT: DO NOT insert probe into stack before turning ON
the analyzer!
1. Place the probe in an area that contains fresh air. This ensures that the
sensors will be properly zeroed during the warm-up cycle.
2. Turn ON the analyzer by pressing the I/O button for at least 1 second, or
until a single beep is heard. Observe that the analyzer’s firmware
version, model and serial numbers are briefly displayed followed by the
Warm Up screen.
3. Wait for the analyzer to count down its 60-second warm-up period, after
which, the instrument will display the Combustion Test HOLD screen.
4. If problems were detected during warm up, the message “ERRORS
DETECTED” is displayed along with a list of those errors. As an
example, the screen below shows that the battery is low. Refer to Error
Messages (page 89) for a listing and possible remedy for the errors
displayed.
NOTE: If the sensors in error are not critical to the
combustion test, then press the RUN/HOLD button to
display the Combustion Test HOLD screen and proceed
with the test.
Figure 5-1. Turning On the Analyzer and Warm-up
0024-9472 Rev 2 41
PCA 3 Manual
5.3. Low Battery Alarm
When the batteries are nearly depleted, an empty battery icon appears in the
upper right corner of the display, and a short beep is sounded every 10 seconds.
After a low battery alarm occurs, the analyzer will continue to operate for only a
few minutes. The amount of operating time that remains depends on many
factors (e.g., pump and keypad backlight being ON or OFF, and the type and
condition of the batteries).
5.4. Locating Sampling Points
Forced Air Furnace Hot Water Tank
Figure 5-3. Hot Water Tank
Figure 5-2. Forced Air Furnace
For atmospheric burner or gravity vented, Domestic hot water tanks with the ‘bell’
forced air heating equipment with a shaped draft diverter can be accurately
clamshell or sectional heat exchanger tested by inserting the probe tip directly
design, test each of the exhaust ports at into the top of the fire tube below the
the top of the heat exchanger. The probe diverter.
should be inserted back into each of the
exhaust ports to obtain a flue gas sample,
before any dilution air is mixed in.
42 0024-9472 Rev 2
PCA 3 Manual
80% Efficiency Fan
90% Efficiency Atmospheric/Gravity
Assist or Power Vented
Condensing Furnace Vented Boiler
Furnace
Figure 5-6.
Figure 5-5. 80%
Figure 5-4. 90%
Atmospheric/Gravity
Efficiency Condensing Efficiency Fan Assist or
Vented Boiler
Furnace Power Vented Furnace
Condensing Combustion testing of fan Boilers, which have a ‘bell’
furnaces/boilers can be assist or power vented, shaped draft diverter on
tested through a hole furnaces/boilers should be top, should be tested
drilled in the plastic vent done through a hole drilled directly below the diverter
pipe (when allowed by the in the vent immediately through a hole drilled in the
manufacturer or local above the inducer fan. vent connector.
authority of jurisdiction) or
taken from the exhaust
termination.
5.5. Performing a Combustion Test
Ensure that the following items have been completed.
Turn ON analyzer and allow it to warm up (see page 41)
•
• Select fuel being burned (see page 23)
Inset probe into stack (see page 42)
•
• If necessary, insert the optional primary air thermocouple into the
combustion air stream of the burners that use an outside source of
combustion air.
Next, proceed with the combustion test as described below.
0024-9472 Rev 2 43
PCA 3 Manual
Figure 5-7. Performing a Combustion Test
1. Press the RUN/HOLD button to start the test. You should hear the
pump start running and see the word RUN appear at the top of the
Combustion Test screen.
The following indicators appear in the
Sensor Indicators:
sensor’s data field depending on certain conditions:
(* * *) Sensor that is not calibrated or installed.
(XXX) Sensor over range
( ) The calculated data cannot be displayed
- - -
because the measured data necessary to
make the calculation is out of range (i.e.,
oxygen level above 16%).
2. Use the up and down arrow keys to scroll to the T-STK reading. Loosen
the thumbscrew on the probe stop and move the probe in and out of the
stack until the stack’s core temperature (hot spot) is located as indicated
by the highest T-STK reading; then tighten the thumbscrew to prevent
further probe movement. Locating the highest stack temperature is very
important for accurate efficiency calculations.
3. You can now begin burner service procedures. The analyzer readings
will change quickly to show changes in burner performance.
CAUTION: Position the water trap with its gas flow arrow
pointing upward. Do not let water condensate go above the
tip of the riser tube. The sensors could be damaged if
water would enter the analyzer. Empty the water trap after
every combustion test. See Emptying the Water Trap on
pages 49 and 83.
4. Pressing the RUN/HOLD button holds all readings, stops the pump and
displays the Combustion Test HOLD screen. Press the ▲▼ buttons to
scroll through all test values. Pressing RUN/HOLD again restarts the
pump and resumes testing.
44 0024-9472 Rev 2
PCA 3 Manual
List of Combustion Test Data
Display Name Description of Measurement or Calculation
O % Oxygen
2
CO Carbon Monoxide (1)
EFF Combustion Efficiency
CO % Carbon Dioxide
2
T-STK Stack Temperature
T-AIR Primary/Ambient Air Temperature as measured either internally or by
an optional external thermocouple plugged into the analyzer’s T-AIR
connector.
EA % Excess Air
CO(#) Carbon Monoxide ppm level referenced to a % of oxygen (2)
NO Nitric Oxide (1)
NO Nitrogen Dioxide (1)
2
NOx Oxides of Nitrogen (NO and NO combined) (1)
2
SO Sulfur Dioxide (1)
2
NO(#) Nitric Oxide ppm level referenced to a % of oxygen (2)
NO (#) Nitrogen Dioxide ppm level referenced to a % of oxygen (2)
2
NOx(#) Oxides of Nitrogen ppm level referenced to a % of oxygen (2)
SO (#) Sulfur Dioxide ppm level referenced to a % of oxygen (2)
2
NO-T Nitric Oxide Sensor Temperature (3)
(1) See page 27 for pollution unit of measure selected.
(2) The “#” represents the selected oxygen reference level of between 0 and 15% (see
page 29).
(3) Only shown if selected in the Run/Hold Format menu (see page 37).
NOTE: Stars (* * *) appear in measurement and calculation fields of
sensors that are not installed.
5.6. Selecting a Pressure Label
The pressure measurement can be labeled with types including, differential
across heat exchanger, draft reading, and differential pressure. See page 47 for
pressure and draft measurement procedures. Label data as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight PRESSURE, and then press ENT to
display the PRESSURE MENU.
3. Use the ◄► buttons to scroll through the available pressure label
types. The current label is used, saved, and printed with the record.
0024-9472 Rev 2 45
PCA 3 Manual
Figure 5-8. Selecting a Pressure Label
5.7. Temperature Label Selection and Measurement
The difference in temperature between two areas can be measured by using the
analyzer’s two temperature channels and the Temperature screen. By using the
T-Air channel as a reference, the temperature applied to the T-Stack channel will
be displayed on the Temperature Measurement screen as differential
temperature between the two channels. Additionally, the temperature
measurement can be labeled. Label types include differential temperature
across heat exchange or differential temperature. Perform a differential
temperature measurement as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight TEMPERATURE, and then press ENT
to display the TEMPERATURE MEASUREMENT screen.
3. Install thermocouples in both temperature channel connectors.
4. Before taking a measurement, the temperature channels may need to
be zeroed, if not already displaying zero with both thermocouples
exposed to the same condition. Press ZERO (F2) if needed.
5. To label the data use the ◄► buttons to scroll through the available
temperature label types. The current label will be used, saved, and
printed with the record.
6. Press ENT to save the selection and re-display the
TEMPERATURE MENU or ESC to exit back to the Main Menu.
Figure 5-9. Performing a Temperature Measurement
46 0024-9472 Rev 2
PCA 3 Manual
5.8. Performing a Draft / Pressure Measurement
The difference in pressure ( ∆P) between two areas can be measured by using
the analyzer’s two pressure ports and the PRESSURE screen. By using the –∆P
port as the reference, the pressure applied to the + ∆P port will be displayed on
the PRESSURE screen as the differential pressure between the two ports.
Perform a draft / pressure measurement as follows.
1. Turn ON the analyzer and allow it to complete its warm-up cycle (see
page 9, 16, and 20).
NOTE: Refer to Pressure Units Selection on page 26 for
information on selecting the pressure units of measure.
2. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
3. Use the ▲▼ buttons to highlight PRESSURE, and then press ENT to
display the PRESSURE screen.
4. Before taking a measurement, the pressure sensor may need to be re-
zeroed if it is not already displaying zero with both pressure ports open
to the atmosphere. If necessary, zero the pressure sensor as follows.
Press the button.
• ZERO (F2)
Disconnect any hoses connected to the +∆P and –∆P
•
ports, and then press ENT to zero the pressure sensor.
• Reconnect any hoses. When measuring draft, simply
leave the -∆P port open to the atmosphere and connect
the probe’s draft hose to the +∆P port (see Figure 5-12).
Figure 5-10. Performing a Draft/Pressure Measurement
5. Do one of the following to measure draft or differential pressure:
• To measure draft, simply insert the probe into the stack
and observe the draft reading on the PRESSURE screen.
To measure differential pressure, connect two sampling
•
hoses to the +∆P and –∆P ports, and place the open end
of each hose into the areas being measured. The
0024-9472 Rev 2 47
PCA 3 Manual
differential pressure between the two areas is now
displayed on the PRESSURE screen. If the pressure at
the +∆P port is higher than the –∆P port, then the pressure
reading will be positive. If it is lower, then the reading will
be negative.
Figure 5-12. Draft/Pressure
Figure 5-11. Pressure Zero Screen
Measurement Hose Connections
5.9. Saving Test Data
Up to 500 individual sets (“snap shots”) of combustion test, pressure, or
temperature data can be saved in memory, which can later be recalled for
viewing from the Memory Directory (see Recalling Combustion Test Data section
on page 54).
1. First display the screen that contains the data to be saved. In the
example below, all data associated with the Combustion Test RUN
screen will be saved.
2. Press the SAVE (F3) button to save the test data in the next available
memory location.
NOTE: When memory is full, the next reading will not be
saved until space is made available by clearing the data
(see Clearing Memory section on page 56).
Figure 5-13. Saving Test Data
48 0024-9472 Rev 2
PCA 3 Manual
5.10. Ending a Combustion Test
WARNING: Burn Hazard. Do not touch the probe after removing it
from the stack. Allow the probe to cool before handling (about 5
minutes).
1. Remove probe from stack.
2. Allow the pump to run until all combustion gases have been flushed
from the analyzer with fresh air as indicated by the O reading returning
2
to 20.9%.
5.11. Emptying the Water Trap
The Water Trap / Filter Assembly removes stack gas condensate, and also
prevents soot from contaminating the internal components of the analyzer.
IMPORTANT: Use the Water Trap / Filter Assembly in a vertical
position with the gas flow arrow pointing up as shown in the
illustration below.
1. Empty the water trap chamber after each combustion test, or stop the
test and empty the chamber if the liquid condensate level approaches
the tip of the riser tube.
2. To empty the trap, first pull apart the two halves of the Water Trap using
a slight twisting motion; empty the water trap chamber; and then
reassemble the trap.
3. After each combustion test, also check the Water Trap’s filter element. If
it looks dirty, replace the filter (see Replacing the Filter Element on page
83).
0024-9472 Rev 2 49
PCA 3 Manual
Figure 5-14. Emptying the Water Trap
5.12. Data Logging
5.12.1. Overview
When the logging function is activated, up to 500 combustion test records will be
automatically stored in memory at a preset interval over a predetermined
duration.
The maximum duration that data can be collected is determined by the interval.
For example, if the interval is set to 10 seconds, then the maximum selectable
length of time in which data can be collected to fill 500 memory locations would
be 1 hour (500 x 10 seconds = 5000 seconds or 83 minutes). If the operator
chooses a duration that is longer than possible for the interval chosen, then the
analyzer automatically selects the highest duration possible for the selected
interval.
The stored logged data can either be viewed on the display using the analyzer’s
memory function (see Memory Functions on page 54), or downloaded to a
®
personal computer using the supplied Fyrite User Software and USB cable.
Downloaded data is stored on the computer’s hard drive as a comma separated
value ASCII text file with a CSV extension, which can be opened by most
spreadsheet programs for analysis.
50 0024-9472 Rev 2
PCA 3 Manual
Logging Intervals and Durations
Logging Intervals Logging Durations
1 second 5 minutes
5 second 10 minutes
10 second 15 minutes
15 second 30 minutes
30 second 1 hour
1 minute 2 hours
2 minutes 5 hours
5 minutes 10 hours
10 minutes 24 hours
48 hours
5.12.2. Turning On Data Logging
IMPORTANT: Before turning ON data logging and starting the data
logging process, the analyzer should already be set up to perform a
combustion test (see Performing a Combustion Test section on
page 43).
Turn ON data logging as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight LOGGING, and then press ENT to
display the LOGGING screen.
NOTE: The amount of memory available for storing new
data is displayed in the LOGGING screen (maximum of
500 locations). If previous logging sessions are stored, and
additional memory is needed, clear the logging memory
(see Clearing Memory section on page 56).
NOTE: At no time will new logging data over write old
data. Logging will stop when available memory is filled.
4. Use the ▲▼ buttons to highlight YES (turn logging ON).
5. Press ENT to make the selection and display the LOGGING INTERVAL
screen.
0024-9472 Rev 2 51
PCA 3 Manual
Figure 5-15. Turning On Data Logging
5.12.3. Setting the Logging Interval and Duration
The logging interval is the length of time between measurements, while the
logging duration is the time allocated to the logging process. Set the logging
interval and duration as follows.
NOTE: If the duration is set for more than 10 hours, we recommend
using the optional AC power adapter to power the analyzer.
1. As soon as logging is turned ON (see Logging Selection on page 31),
the LOGGING INTERVAL screen appears.
2. Use the ▲▼ buttons to highlight the desired interval, and then press
ENT to make the selection and display the LOGGING DURATION
screen. In this example, 1 minute has been selected.
3. Use the ▲▼ buttons to highlight the desired duration, and then press
ENT to make the selection and display the LOGGING SUMMARY
screen. In this example, 2 hours has been selected.
NOTE: If the selected duration requires more memory
than is currently available, the analyzer will automatically
select the longest duration possible for the selected
interval.
Figure 5-16. Setting the Logging Interval
52 0024-9472 Rev 2
PCA 3 Manual
5.12.4. Starting the Data Logging Process
After turning ON data logging and setting the interval and duration, the analyzer
will pause at the LOGGING SUMMARY screen, where the currently selected
interval and duration time periods are displayed.
Press to start the combustion test and logging process. At this time the
ENT
Combustion Test LOG screen will appear, indicating that the analyzer is now
performing a combustion test and the data is being stored in memory.
NOTE: At the top of the screen, the memory location where the
current log entry is being saved is displayed.
NOTE: If there were previously stored log entries, each new logging
session will start to store data in the next available memory location.
NOTE: At no time will new data overwrite old data. Logging will
stop when available memory is filled.
Figure 5-17. Starting the Data Logging Process
5.12.5. Ending the Data Logging Process
Data logging will stop and the pump will turn OFF after the prescribed duration,
or after all 500 memory locations are filled.
To exit the Combustion Test LOG screen, press the button twice to
RUN/HOLD
display the Combustion Test HOLD screen.
NOTE: To end the logging process at any time, press the ESC
button.
0024-9472 Rev 2 53
PCA 3 Manual
NOTE: If the logging process was stopped by pressing ESC, it
cannot be resumed without first turning logging back ON.
5.13. Memory Functions
5.13.1. Overview
There are two memory banks, each containing 500 memory locations. The first
bank is used to store combustion test data as described in Saving Test Data
(page 48), while the second bank is used to store logged combustion test data as
described in Data Logging (page 50). Each bank is independent of each other,
and cannot share data or be combined.
Individual memory locations in each memory bank can be recalled for viewing on
the display or printed, or the entire contents of each memory bank can be
individually downloaded to a computer and viewed in a spreadsheet program for
analysis. These functions are explained in the sections that follow.
NOTE: When displaying the contents of either memory bank, the
operator can quickly page through the screens by pressing the
PAGE– (F1) and PAGE+ (F3) buttons. Or move to the first or last
memory location by pressing the ◄► buttons, respectively.
5.13.2. Recalling Combustion Test Data
Recall individual combustion test data records as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight MEMORY, and then press ENT to
display the MEMORY MENU.
3. Use the ▲▼ buttons to highlight MEMORY DIRECTORY, and then
press ENT to display the MEMORY DIRECTORY screen.
4. Use the ▲▼ buttons to highlight the desired memory location to be
recalled. Each memory location is identified by the date and time at
which data was saved. The word EMPTY signifies that the memory
location does not contain data.
5. Press ENT to display the data contained in the selected memory
location.
NOTE: The top line of the recalled combustion test data screen
shows the memory location being viewed. In this example, “MEM: 1”
is being displayed.
54 0024-9472 Rev 2
PCA 3 Manual
NOTE: The recalled combustion test data can be printed by
pressing the PRINT (F1) button. Refer to Printing Test Data on
page 59.
Figure 5-18. Recalling Combustion Test Data
5.13.3. Recalling Logged Test Data
Recall individual logged combustion test data records as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight MEMORY, and then press ENT to
display the MEMORY MENU.
3. Use the ▲▼ buttons to highlight LOGGING DIRECTORY, and then
press ENT to display the LOGGING DIRECTORY screen.
4. Use the ▲▼ buttons to highlight the desired data logging location to be
recalled. Each logging location is identified by the date and time at
which the data was saved. The word EMPTY signifies that the logging
location does not contain data.
5. Press ENT to display the logging data contained in the selected memory
location.
NOTE: The top line of the log test data screen shows the log
location being viewed. In this example, “Log: 1” is being displayed.
NOTE: The recalled logged test data can be printed by pressing the
PRINT (F1) button. Refer to Printing Test Data on page 59.
0024-9472 Rev 2 55
PCA 3 Manual
Figure 5-19. Recalling Logged Test Data
5.13.4. Clearing Memory
When all memory locations used to store individual combustion test records have
been filled, the next combustion test record saved will not overwrite the oldest.
When all logging memory locations in the logging directory are full, they must be
manually cleared in order to store new data. At no time will the logging process
overwrite older data. Logging will stop when available memory is filled. Each
memory bank can be individually cleared, or all memory locations in both banks
can be cleared simultaneously. Do the following to clear memory:
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight MEMORY, and then press ENT to
display the MEMORY MENU.
Figure 5-20. Clearing Memory
3. Use the ▲▼ buttons to highlight one of the following:
• CLEAR MEMORY DATA clears only the individual saved
combustion test records.
• CLEAR LOGGING DATA clears only the combustion test records
that were saved during the logging process.
• CLEAR ALL MEMORY clears all memory locations in both memory
banks.
4. Press ENT to display the CLEAR MEMORY, LOGGING, or ALL DATA
conformation screen. Highlight YES to confirm that memory is to be
cleared, and then press ENT to actually clear memory as evidenced by
the display of the “WAIT Erasing memory” screen.
56 0024-9472 Rev 2
PCA 3 Manual
5.14. Downloading Stored Data to a Computer
The combustion test data that was stored in either the analyzer’s Memory
Directory or Logging Directory (see Memory Functions on page 54), can be
®
downloaded to a computer using the Fyrite User Software and USB cable that
are supplied with the analyzer.
The downloaded data is stored on the computer’s hard drive – or removable
media of the operator’s choosing – as a comma-separated-value ASCII text file
with a CSV extension. This type of file can be opened by most spreadsheet
programs for analysis.
®
For information on the installation of the Fyrite User Software, refer to the
®
installation document (provided in .PDF format) on the software disk. The Fyrite
®
User Software installation manual may also be downloaded from the PCA 3
page of the Bacharach website (www.MyBacharach.com). Refer to manual
0024-9465. Minimum computer requirements for installation and use of the
®
Fyrite User Software are shown below.
®
Minimum Computer Requirements for Fyrite User Software
Category Minimum Requirement
Operating System Windows XP or higher
Drive CD ROM drive
Access Port USB port (USB 1.1, 2.0, or 3.0)
®
Software: 205 MB for the Fyrite User Software
Hard Drive Space
Data: Additional 350 KB for each downloaded file
5.15. Importing Saved Data into a Spreadsheet
Data that was recovered and saved as an ASCII text file with a “CSV” extension,
as described above, can easily be opened for viewing in most spreadsheet
programs by simply double-clicking the filename. For example: double-clicking
the filename Customer XYZ.csv should automatically open the spreadsheet
program and display the contents of the file.
If the spreadsheet program does not recognize the “CSV” file extension, then
refer to the spreadsheet’s documentation for information on how to manually
import comma-delimited text files.
The table below contains listings and descriptions of the 32 data fields that are
downloaded with each data record. Note that for sensors that are not installed, all
related data fields for those sensors will be marked as “***”. For example, if the
analyzer does not contain an SO sensor, then “***” will appear in data fields 23
2
and 24.
0024-9472 Rev 2 57
PCA 3 Manual
Downloaded Data Fields
Field Column Name Data Name or Value
1 Date Date of Test in mm/dd/yy
2 Time Time of Test in 12 hour format (AM/PM)
3-5 Test ID Information Input by User
6 Fuel Fuel Name
7 %O Oxygen Level
2
8 CO Carbon Monoxide Level (see field 30)
9 ppm CO(O ) Carbon Monoxide Level referenced to a % of O (see field 25)
2 2
10 % Efficiency Combustion Efficiency
11 % CO Carbon Dioxide Level
2
12 %EA Excess Air
13 TStk Stack Temperature
14 TAir Primary/Ambient Air Temperature
15 Delta-T Differential Temperature Value
16 TempUnit Temperature Units
17 NO Nitric Oxide Level (see field 30)
18 ppm NO(O ) Nitric Oxide Level referenced to a % of O (see field 26)
2 2
19 NO Nitrogen Dioxide Level (see field 30)
2
20 ppm NO (O ) Nitrogen Dioxide Level referenced to a % of O (see field 26)
2 2 2
21 NO Nitrogen Oxides (NO+NO ) (see field 30)
x 2
(see field 26)
22 ppm NO (O ) Nitrogen Oxides Level as referenced to a % of O
x 2 2
23 SO Sulfur Dioxide (see field 30)
2
24 ppm SO (O ) Sulfur Dioxide Level referenced to a % of O (see field 27)
2 2 2
(1)
25 CO_O Ref %O Reference for CO
2 2
(1)
26 NO _O Ref %O Reference for NOx
x 2 2
(1)
27 SO _O Ref %O Reference for SO
2 2 2 2
28 Pressure Pressure (Draft) Value
(2)
29 Pressure Units Pressure Units of Measure
(3)
30 Pollution Units Pollution Units of Measure
31 PS Voltage Power Supply Voltage
32 NO Temp Nitric Oxide Sensor Temperature
(1) %O reference as selected (refer to O2 Reference Setup on page 29)
2
(2) Pressure units as selected (refer to Pressure Units Selection on page 26)
(3) Pollution units as selected (refer to Pollution Units Selection on page 27)
58 0024-9472 Rev 2
PCA 3 Manual
5.16. Printing Test Data
Combustion or pressure data that is currently being displayed can be sent to a
printer using IrDA protocol as described below.
Data that is stored in memory can also be printed by first displaying the stored
test data as described in the Recalling Combustion Test Data section (page 54)
and the Recalling Logged Test Data section (page 55).
In addition to printing combustion and pressure data, the contents of any screen
that shows the label “PRINT” above the F1 button can be printed. For example,
the information shown in the DIAGNOSTIC screen can be printed.
1. Turn ON printer. Refer to the printer’s instruction manual for detailed
operating information. If not already done, set up the printer for:
8 bit
•
No parity
•
9600 baud
•
IrDA is set to IrDA-SIR
•
• DTR handshaking.
2. Align the printer with the top of the analyzer as shown in Figure 5-21.
3. Press the PRINT (F1) button to begin printing.
The printout shown in Figure 5-21 shows typical combustion test results of an
analyzer containing O , CO , NO, and NO sensors. Note that since the SO
2 LOW 2 2
sensor is not installed, three stars (* * *) appear in their data fields.
NOTE: Three Stars (***) appear in the data field of sensors that are
not calibrated or installed, and in the data fields of that sensor’s
related calculated values.
NOTE: Three hyphens or dashes (---) indicate that the calculated
data cannot be displayed because the measured data necessary to
make the calculation is out of range (i.e., oxygen level above 16%).
NOTE: Three Xs (i.e., X X X) indicate the sensor is over range.
0024-9472 Rev 2 59
PCA 3 Manual
Figure 5-21. Printer Alignment and Sample Printout
60 0024-9472 Rev 2
PCA 3 Manual
5.17. Turning Off the Analyzer and Purging
Turn OFF the analyzer by pressing the I/O button for at least 2 seconds, or until
two beeps are heard. The unit will count down 5 seconds before shutting down,
giving the operator an opportunity to keep the analyzer turned ON by pressing
the button.
RUN/HOLD
®
If the PCA 3 was not purged with fresh air (as described in Ending a Combustion
Test on page 49), then the analyzer may remain ON with its pump running and
display the message “PURGING SENSORS” as the result of combustion gases
still being present inside the analyzer. At this time the operator should ensure
that the probe is removed from the stack, allowing the analyzer to purge itself
with fresh air. The 5-second shutdown sequence will not begin until the gas
levels inside the analyzer drop below predetermined levels.
NOTE: Although not recommended, the purging process can be
bypassed by pressing the I/O button a second time.
0024-9472 Rev 2 61
PCA 3 Manual
CHAPTER 6. CALIBRATION
IMPORTANT: Before performing any calibration procedure ensure
that new batteries are installed or use the optional AC power
adapter. Also ensure that the analyzer is at room temperature and
will be sampling fresh air when turned ON.
6.1. B-Smart Sensors
® ®
The PCA 3 uses Bacharach’s B-Smart Sensor technology for CO , CO ,
LOW HIGH
®
NO, NO , and SO . The B-Smart Sensor is marked with a 10- or 14-digit
2 2
calibration code that can be entered in the instrument’s calibration screen via the
®
keypad, or with the Fyrite User Software.
®
Benefits of the B-Smart Sensors include the following:
New sensors can be installed without applying gas for calibration.
•
• Sensors can be pre-calibrated and installed when needed
Sensors can be moved from one analyzer to another.
•
• The analyzer’s diagnostics screen shows what sensors are installed,
and their current end-of-life condition (Good, Low, Bad).
®
B-Smart sensors should be calibrated by an authorized Bacharach Service
Center (see Service Centers on page 95) every 6 months to 1 year to assure that
®
the analyzer continues to meet its published accuracy specifications. B-Smart
sensors, however, can be calibrated in the field if your facility has the necessary
equipment and qualified personnel to perform the procedures described in the
following sections.
6.2. Starting a Calibration
Start any calibration procedure by doing the following.
1. Place the probe in an area of fresh air, turn ON the analyzer, allowing it
to cycle through its 60 second warm-up period. During warm-up, the
analyzer’s operation is checked and the sensors are set to the following
ambient conditions:
Oxygen sensor spanned to 20.9%
•
• All gas sensors are zeroed
The pressure sensor is zeroed
•
2. Any errors detected during warm-up will be listed on the display
immediately following warm-up. Correct any errors before proceeding.
Refer to page 89 (Error Messages) for a listing of error messages and
their meaning.
3. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
4. Use the ▲▼ buttons to highlight CALIBRATION, and then press ENT to
display the CALIBRATION PASSWORD screen.
62 0024-9472 Rev 2
PCA 3 Manual
5. Before calibration can begin a 4-character alphanumeric password must
be entered. Use the ▲▼ and ◄► buttons to enter the password. (The
default password is 1111.)
6. Press the ENT key to accept the password. If the correct password was
entered the CALIBRATION LIST Menu will be displayed.
7. Use the ▲▼ buttons to highlight the sensor to be calibrated, and then
perform the calibration procedure for that sensor as described in the
following sections.
Figure 6-1. Starting a Calibration
®
6.3. B-Smart Sensor Replacement and Calibration
®
To replace a B-Smart sensor and calibrate, do the following.
1. Enter the CALIBRATION MENU.
®
2. Use the ▲▼ buttons to select B-SMART . Press ENTER to display the
®
B-Smart code screen.
3. Use the ▲▼ buttons to enter the 10- or 14-digit alphanumeric code
®
supplied with the B-Smart sensor. Use the ◄► buttons to move the
cursor across the screen. Press ENTER.
®
NOTE: There are two B-Smart sensor code lengths: 10 digits for
CO , NO, NO , and SO , and 14 digits for CO .
HIGH 2 2 LOW
NOTE: If the correct code was entered, the analyzer accepts it and
returns to the CALIBRATION MENU. If an incorrect code was
entered, the screen will display “Invalid Code”. Check to make sure
the correct code has been entered. If the problem persist, contact
your nearest Bacharach Service Center.
®
NOTE: To save time, B-Smart codes can be entered using the
Fyrite User Software.
0024-9472 Rev 2 63
PCA 3 Manual
Bacharach also offers a convenient exchange program that allows the customer
to regularly send in old sensors and, as scheduled, receive calibrated
replacements. Contact Bacharach customer service for more details about the
exchange program.
6.4. Pressure Sensor Calibration
This procedure calibrates the pressure sensor to a known pressure value.
6.4.1. Material Required
Bellows
•
• Manometer:
Range: ±8 in. of water column (±20 mb)
Accuracy: ±0.01 in. of water column (±0.025 mb)
6.4.2. Procedure
NOTE: The unit of measure for pressure is selected in the Pressure
Units Selection section on page 26. In the following procedure, inwc
is selected, but note that any unit of measure can be used for
calibration purposes.
1. Assemble the pressure sensor calibration equipment as shown in
Figure 6-3, but DO NOT connect the analyzer to the calibration
equipment at this time.
2. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen (refer to Calibration on page 62).
3. Use the ▲▼ buttons to highlight Pressure, and then press ENT to
display the CALIBRATE PRESSURE screen.
NOTE: “Measured” is the pressure value currently being
detected by the pressure sensor, while “Applied” is a
known value of pressure that will be applied for calibration
purposes.
4. With both the –∆P and +∆P ports open to the atmosphere, observe that
the current Measured pressure reading should be 0 ±0.01 inwc. If
necessary, zero the pressure sensor (refer to Performing a Draft /
Pressure Measurement on page 47) and repeat Steps 2 thru 4.
5. Connect the hose from the manometer to the +∆P port and apply a
negative pressure to this port by adjusting the bellows for a manometer
reading of –4.00.
6. Use the ▲▼ and ◄► buttons to enter an applied value that exactly
equals the manometer reading.
64 0024-9472 Rev 2
PCA 3 Manual
NOTE: The calibration range is from –6 to –2 inwc (–15 to
-5 mb). An attempt to calibrate outside this range will
cause the message “Applied Value High” (or Low) to
appear at the bottom of the screen.
Figure 6-2. Pressure Sensor Calibration
7. Wait until the measured reading stabilizes, and then press ENT to
calibrate the pressure sensor’s Measured value to that of the applied
value; after which the message “Good Calibration” should briefly appear
followed by the CALIBRATION LIST screen being re-displayed.
8. Remove calibration equipment.
Figure 6-3. Pressure Sensor Calibration Equipment
6.5. T-Stack Calibration
This procedure first zeros and then spans the stack temperature channel to
known temperature values.
0024-9472 Rev 2 65
PCA 3 Manual
The use of an electronic thermocouple simulator is the preferred method of
producing the desired calibration temperatures. Alternatively, containers of ice
water and boiling water can be used.
6.5.1. Material Required
Thermocouple Simulator (K-type):
•
Range: 0 to 600 °F
Accuracy: ±0.5 °F
(Alternatively) Ice Water, Boiling Water, Thermometer
•
6.5.2. TS-Zero Procedure
1. Set thermocouple simulator to room temperature and plug its output into
the T-STACK connector located at the bottom of the analyzer.
Alternatively: Plug the probe’s thermocouple into the T-
STACK connector located at the bottom of the analyzer.
CAUTION: DO NOT attach the probe’s gas hose to the
analyzer’s GAS port or water will be drawn into the
analyzer!
2. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen (refer to Calibration on page 62).
3. Use the ▲▼ buttons to highlight T-Stack, and then press ENT to
display the CALIBRATE TS-ZERO screen.
NOTE: “Measured” is the current temperature reading,
while “Applied” is a known temperature that will be applied
for calibration purposes.
4. Set thermocouple simulator to 32 °F (0 °C), and then use the ▲▼ and
◄► buttons to enter an applied value that exactly equals the setting of
the simulator.
Submerge probe tip into an ice-water bath
Alternatively:
with a thermometer, wait several minutes, and then use the
▲▼ and ◄► buttons to enter an applied value that exactly
equals the thermometer reading.
NOTE: The calibration range is from 32 to 41 °F (0 to 5
°C). Calibrating outside this range will cause the message
“Applied Value High” (or Low) to appear at the bottom of
the screen.
5. Wait until the measured reading stabilizes, and then press ENT to
calibrate the TS-Zero Measured value to that of the applied value; after
66 0024-9472 Rev 2
PCA 3 Manual
which the message “Good Calibration” should briefly appear followed by
the CALIBRATE TS-SPAN screen.
Figure 6-4. T-Stack Zero and Span
6.5.3. TS-Span Procedure
1. Set thermocouple simulator to 572 °F (300 °C), and then use the ▲▼
and ◄► buttons to enter an applied value that exactly equals the
setting of the simulator.
Alternatively: Submerge probe tip into a container of boiling
water with a thermometer, wait several minutes, and then use
the ▲▼ and ◄► buttons to enter an applied value that exactly
equals the thermometer reading.
NOTE: The calibration range is from 175 to 625 °F (80 to
330 °C). An attempt to calibrate outside this range will
cause the message “Applied Value High” (or Low) to
appear at the bottom of the screen.
2. Wait until the measured reading stabilizes, and then press ENT to
calibrate the TS-Span Measured value to that of the applied value; after
which the message “Good Calibration” should briefly appear followed by
the CALIBRATION LIST screen being re-displayed.
6.6. T-Air Calibration
This procedure first zeros and then spans the ambient temperature channel to
known temperature values.
The use of an electronic thermocouple simulator is the preferred method of
producing the desired calibration temperatures. Alternatively, containers of ice
water and boiling water can be used.
6.6.1. Material Required
• Thermocouple Simulator (K-type)
Range: 0 to 600 °F
Accuracy: ±0.5 °F
• (Alternatively) Ice Water, Boiling Water, Thermometer
6.6.2. TA-Zero Procedure
0024-9472 Rev 2 67
PCA 3 Manual
1. Set thermocouple simulator to room temperature and plug its output into
the T-AIR connector located at the bottom of the analyzer.
Plug the probe’s thermocouple into the T-AIR connector
Alternatively:
located at the bottom of the analyzer. DO NOT attach the probe’s gas
hose to the analyzer’s GAS port; otherwise water will be drawn into the
analyzer!
2. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen (refer to Calibration on page 62).
3. Use the ▲▼ buttons to highlight T-Air, and then press ENT to display
the CALIBRATE TA-ZERO screen.
NOTE: “Measured” is the current temperature reading,
while “Applied” is a known temperature that will be applied
for calibration purposes.
4. Set thermocouple simulator to 32 °F (0 °C), and then use the ▲▼ and
◄► buttons to enter an applied value that exactly equals the setting of
the simulator.
Alternatively: Submerge probe tip into an ice-water bath with a
thermometer, wait several minutes, and then use the ▲▼ and ◄►
buttons to enter an applied value that exactly equals the thermometer
reading.
NOTE: The calibration range is from 32 to 41 °F (0 to
5 °C). An attempt to calibrate outside this range will cause
the message “Applied Value High” (or Low) to appear at
the bottom of the screen.
5. Wait until the measured reading stabilizes, and then press ENT to
calibrate the TA-Zero Measured value to that of the applied value; after
which the message “Good Calibration” should briefly appear followed by
the CALIBRATE TA-SPAN screen.
Figure 6-5. T-Air Zero Procedure
68 0024-9472 Rev 2
PCA 3 Manual
6.6.3. TA-Span Procedure
1. Set thermocouple simulator to 212 °F (100 °C), and then use the ▲▼
and ◄► buttons to enter an applied value that exactly equals the
setting of the simulator.
Alternatively: Submerge probe tip into a container of boiling water
with a thermometer, wait several minutes, and then use the ▲▼ and
◄► buttons to enter an applied value that exactly equals the
thermometer reading.
NOTE: The calibration range is from 194 to 230 °F (90 to
110 °C). An attempt to calibrate outside this range will
cause the message “Bad Calibration Wrong CAL Entry” to
appear in the following step.
2. Wait until the measured reading stabilizes, and then press ENT to
calibrate the TA-Span Measured value to that of the applied value; after
which the message “Good Calibration” should briefly appear followed by
the CALIBRATION LIST screen being re-displayed.
Figure 6-6. T-Air Span Procedure
6.7. CO Sensor Calibration
LOW
NOTE: The CO sensor also measures H for the purpose of
LOW 2
compensating the CO reading for the presence of H in the gas
2
sample. This procedure first spans the CO sensor and,
LOW
optionally, spans the H part of the sensor to known gas levels.
2
The user has the option to perform a CO only calibration or a CO/H calibration
2
®
using the PCA 3 soft keys.
6.7.1. Material Required
Calibration Kit (P/N 0024-7059)
•
Gas Cylinder (500 ppm CO in air, P/N 0024-0492)
•
Gas Cylinder (1,000 ppm CO and 1,000 ppm H in Nitrogen,
• 2
P/N 0024-0794)
0024-9472 Rev 2 69
PCA 3 Manual
6.7.2. Procedure
1. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen (refer to Calibration on page 62).
2. Use the ▲▼ buttons to highlight CO-LO, and then press ENT to display
the CALIBRATE CO screen.
NOTE: “Measured” is the current CO reading, while
“Applied” is a known CO level that will be applied for
calibration purposes.
3. Attach a 500 ppm CO cylinder to the regulator of the calibration fixture
shown in Figure 6-12 on page 76.
4. Use the ▲▼ and ◄► buttons to enter an applied value that exactly
equals the concentration stamped on the CO cylinder.
NOTE: The calibration range is from 9 to 1,500 ppm. An
attempt to calibrate outside this range will cause the
message “Applied Value High” (or Low) to appear at the
bottom of the screen.
Figure 6-7. CO Sensor Calibration
LOW
5. Adjust the regulator for a flow meter indication of approximately 2 SCFH.
Wait until the measured reading stabilizes (approximately 3 minutes),
and then press ENT to calibrate the CO Measured value to that of the
applied value. The message “Good Calibration” should briefly appear
followed by the TEST GAS CO/H screen.
2
NOTE: If the sensor’s output is low, but still usable, then
the message “Good Calibration WARNING Low Sensor”
will appear. The sensor will now be marked as being Low
in the DIAGNOSTICS screen.
70 0024-9472 Rev 2
PCA 3 Manual
NOTE: If the sensor’s output is too low to be usable, then
the message “Bad Calibration Sensor End of Life, Entry
Not Saved” will appear. The sensor will now be marked as
being BAD in the DIAGNOSTICS screen.
6. Turn OFF regulator of calibration fixture and remove the CO cylinder.
7. Attach a combination 1,000 ppm CO and 1,000 ppm H cylinder to the
2
regulator of the calibration fixture, and then use the ▲▼ and ◄►
buttons to enter an applied value that exactly equals the CO
concentration stamped on the cylinder.
8. Adjust the regulator for a flow meter indication of approximately 2 SCFH,
and then press ENT to display the CALIBRATE H screen.
2
NOTE: The calibration range is 400 to 1,500 ppm.
Calibrating outside this range causes the message “Bad
Calibration Wrong CAL Entry” to appear in the next step.
Figure 6-8. CO Sensor Calibration
LOW
9. Wait until the measured reading stabilizes (approximately 3 minutes),
and then press ENT to calibrate the H Measured value to that of the
2
applied value; after which the message “Good Calibration” should briefly
appear followed by the CALIBRATION LIST screen being re-displayed.
NOTE: If the sensor’s output is low, but still usable, then
the message “Good Calibration WARNING Low Sensor”
will appear. The sensor will now be marked as being Low
in the DIAGNOSTICS screen.
NOTE: If the sensor’s output is too low to be usable, then
the message “Bad Calibration Sensor End of Life” will
appear followed by the CALIBRATION LIST screen being
re-displayed. The sensor will now be marked as being BAD
in the DIAGNOSTICS screen.
10. Turn OFF the regulator and remove the gas cylinder.
0024-9472 Rev 2 71
PCA 3 Manual
NOTE: A CO-only calibration can be done by pressing F3 in Step 5
(instead of ENT).
6.8. SO Sensor Calibration
2
This procedure spans the optional sulfur dioxide sensor to a known gas level.
6.8.1. Material Required
• Calibration Kit ( P/N 0024-7059)
• Gas Cylinder (100 ppm SO in Nitrogen, P/N 0024-1158)
2
6.8.2. Procedure
1. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen (refer to Calibration on page 62).
2. Use the ▲▼ buttons to highlight SO , and then press ENT to display
2
the CALIBRATE SO screen.
2
NOTE: “Measured” is the current SO2 reading, while
“Applied” is a known SO level that will be applied for
2
calibration purposes.
3. Attach a 100 ppm SO cylinder to the regulator of the calibration fixture
2
shown in Figure 6-12 on page 76.
4. Use the ▲▼ and ◄► buttons to enter an applied value that exactly
equals the concentration stamped on the SO cylinder.
2
NOTE: The calibration range is from 9 to 1,800 ppm. An
attempt to calibrate outside this range will cause the
message “Bad Calibration Wrong CAL Entry” to appear in
the following step.
Figure 6-9. SO Sensor Calibration
2
5. Adjust regulator for a flow meter indication of approximately 2 SCFH.
Wait until the measured reading stabilizes (approximately 3 minutes),
72 0024-9472 Rev 2
PCA 3 Manual
and then press ENT to calibrate the SO Measured value to that of the
2
applied value. The message “Good Calibration” should briefly appear
followed by the CALIBRATION LIST screen being re-displayed.
NOTE: If the sensor’s output is low, but still usable, then
the message “Good Calibration WARNING Low Sensor”
will appear. The sensor will now be marked as being Low
in the DIAGNOSTICS screen.
NOTE: If the sensor’s output is too low to be usable, then
the message “Bad Calibration Sensor End of Life, Entry
Not Saved” will appear followed by the CALIBRATION
LIST screen being re-displayed. The sensor will now be
marked as being BAD in the DIAGNOSTICS screen.
6. Turn OFF regulator and remove the gas cylinder.
6.9. NO Sensor Calibration
This procedure spans the optional nitric oxide sensor to a known gas level.
6.9.1. Material Required
Calibration Kit (P/N 0024-7059)
•
• Gas Cylinder ( 250 ppm NO in Nitrogen, P/N 0024-1156)
6.9.2. Procedure
1. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen (refer to Calibration on page 62).
2. Use the ▲▼ buttons to highlight NO, and then press ENT to display the
CALIBRATE NO screen.
NOTE: “Measured” is the current NO reading, while
“Applied” is a known NO level that will be applied for
calibration purposes.
3. Attach a 250 ppm NO cylinder to the regulator of the calibration fixture
shown in Figure 6-12 on page 76.
4. Use the ▲▼ and ◄► buttons to enter an applied value that exactly
equals the concentration stamped on the NO cylinder.
NOTE: The calibration range is from 9 to 900 ppm. An
attempt to calibrate outside this range will cause the
message “Bad Calibration Wrong CAL Entry” to appear in
the following step.
0024-9472 Rev 2 73
PCA 3 Manual
Figure 6-10. NO Sensor Calibration
5. Adjust regulator for a flow meter indication of approximately 2 SCFH.
Wait until the measured reading stabilizes (approximately 3 minutes),
and then press ENT to calibrate the NO Measured value to that of the
applied value. The message “Good Calibration” should briefly appear
followed by the CALIBRATION LIST screen being re-displayed.
NOTE: If the sensor’s output is low, but still usable, then
the message “Good Calibration WARNING Low Sensor”
will appear. The sensor will now be marked as being Low
in the DIAGNOSTICS screen.
NOTE: If the sensor’s output is too low to be usable, then
the message “Bad Calibration Sensor End of Life, Entry
Not Saved” will appear followed by the CALIBRATION
LIST screen being re-displayed. The sensor will now be
marked as being BAD in the DIAGNOSTICS screen.
6. Turn OFF regulator and remove the gas cylinder.
6.10. NO Sensor Calibration
2
This procedure spans the optional nitrogen dioxide sensor to a known gas level.
6.10.1. Material Required
Calibration Kit (P/N 0024-7059)
•
• Gas Cylinder (100 ppm NO in Nitrogen, P/N 0024-1157)
2
6.10.2. Procedure
1. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen (refer to Calibration on page 62).
, and then press ENT to display
2. Use the ▲▼ buttons to highlight NO
2
the CALIBRATE NO screen.
2
74 0024-9472 Rev 2
PCA 3 Manual
NOTE: “Measured” is the current NO reading, while
2
“Applied” is a known NO level that will be applied for
2
calibration purposes.
3. Attach a 100 ppm NO cylinder to the regulator of the calibration fixture
2
shown in Figure 6-12 on page 76.
4. Use the ▲▼ and ◄► buttons to enter an applied value that exactly
equals the concentration stamped on the NO cylinder.
2
NOTE: The calibration range is from 9 to 150 ppm. An
attempt to calibrate outside this range will cause the
message “Bad Calibration Wrong CAL Entry” to appear in
the following step.
Figure 6-11. NO Sensor Calibration
2
5. Adjust regulator for a flow meter indication of approximately 2 SCFH.
Wait until the measured reading stabilizes (approximately 3 minutes),
and then press ENT to calibrate the NO measured value to that of the
2
applied value. The message “Good Calibration” should briefly appear
followed by the CALIBRATION LIST screen being re-displayed.
NOTE: If the sensor’s output is low, but still usable, then
the message “Good Calibration WARNING Low Sensor”
will appear. The sensor will now be marked as being Low
in the DIAGNOSTICS screen.
NOTE: If the sensor’s output is too low to be usable, then
the message “Bad Calibration Sensor End of Life, Entry
Not Saved” will appear followed by the CALIBRATION
LIST screen being re-displayed. The sensor will now be
marked as being BAD in the DIAGNOSTICS screen.
6. Turn OFF regulator and remove the gas cylinder.
0024-9472 Rev 2 75
PCA 3 Manual
6.11. CO Sensor Calibration
HIGH
This procedure spans the optional carbon monoxide high sensor (4,001 to
20,000 ppm) to a known gas level.
6.11.1. Material Required
Calibration Kit (P/N 0024-7059)
•
• Gas Cylinder (4,000 ppm CO in Air, P/N 0024-1155)
Figure 6-12. Gas Sensor Calibration Equipment
6.11.2. Procedure
1. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen (refer to Calibration on page 62).
2. Use the ▲▼ buttons to highlight CO-HI, and then press ENT to display
the CALIBRATE CO-HI screen.
NOTE: “Measured” is the current CO reading, while
“Applied” is a known CO level that will be applied for
calibration purposes.
3. Attach a 4,000 ppm CO cylinder to the regulator of the calibration fixture
shown in Figure 6-12 on page 76.
4. Use the ▲▼ and ◄► buttons to enter an applied value that exactly
equals the concentration on the CO cylinder.
76 0024-9472 Rev 2
PCA 3 Manual
NOTE: The calibration range is from 250 to 11,000 ppm.
An attempt to calibrate outside this range will cause the
message “Bad Calibration Wrong CAL Entry” to appear in
the following step.
Figure 6-13. CO Sensor Calibration
HIGH
5. Adjust regulator for a flow meter indication of approximately 2 SCFH.
Wait until the measured reading stabilizes (approximately 3 minutes),
and then press ENT to calibrate the CO Measured value to that of the
applied value. The message “Good Calibration” should briefly appear
followed by the CALIBRATION LIST screen being re-displayed.
NOTE: If the sensor’s output is low, but still usable, then
the message “Good Calibration WARNING Low Sensor”
will appear. The sensor will now be marked as being Low
in the DIAGNOSTICS screen.
If the sensor’s output is too low to be usable, then the
message “Bad Calibration Sensor End of Life, Entry Not
Saved” will appear followed by the CALIBRATION LIST
screen being re-displayed. The sensor will now be marked
as being BAD in the DIAGNOSTICS screen.
6. Turn OFF regulator and remove the gas cylinder.
0024-9472 Rev 2 77
PCA 3 Manual
CHAPTER 7. MAINTENANCE
7.1. Overview
®
Customer maintenance of the PCA 3 is limited to the following:
• Battery replacement (see page 20)
• Sensor re-calibration (see page 62)
Water trap / filter assembly maintenance (see pages 49 and 83)
•
• Sensor replacement (see page 84)
Cleaning the probe (see page 87).
•
All other maintenance should be performed by an authorized Bacharach Service
Center. Refer to Service Centers on page 95.
®
7.2. PCA 3 Disassembly
The following procedure describes and illustrates how to disassemble the
analyzer.
7.2.1. Tools Required
• Medium Phillips Screwdriver
7.2.2. Procedure
1. Unplug all thermocouples from bottom of analyzer.
2. Remove battery cover and then remove batteries.
NOTE: In Step 3, if the sensors are not being replaced, leave the
tubing connected to each sensor’s gas cap, being careful not to put
unnecessary strain on the tubing during the disassembly process.
3. Pull off sensor retainer; and then unplug all sensors.
4. Lay analyzer face down on a padded work surface; and then using a
medium Phillips screwdriver, remove the unit’s four rear-case screws.
5. Lift rear case from analyzer and set aside.
6. Unplug electrical connectors J8, J9, and J14 from printed circuit board.
CAUTION: In Step 7, note that there is tubing connected between
the bottom hose connector plate and the pressure sensor on the
printed circuit board. Do not put unnecessary strain on this tubing
during the disassembly process.
7. Carefully lift battery and pump chassis, along with the bottom hose
connector plate, from analyzer.
8. Lift printed circuit board from analyzer.
78 0024-9472 Rev 2
PCA 3 Manual
Figure 7-1. Removing the Rear Case
Model Number and Corresponding Sensor Configuration
Sensor Position
225 235 245 255 265 275 285
O O O O O O O
#1 2 2 2 2 2 2 2
CO CO CO CO CO CO CO
#2 LOW LOW LOW LOW LOW LOW LOW
Empty Empty Empty SO NO SO CO
#3 2 2 2 HIGH
Empty NO CO Empty NO NO NO
#4 HIGH
0024-9472 Rev 2 79
PCA 3 Manual
Figure 7-2. Inside View with Rear Case Removed
80 0024-9472 Rev 2
PCA 3 Manual
Figure 7-3. Tubing Connections - Main
Figure 7-4. Tubing Connections - Sensors
0024-9472 Rev 2 81
PCA 3 Manual
Figure 7-5. Wire Routing, Connectors, and Solder Connections
82 0024-9472 Rev 2
PCA 3 Manual
7.3. Emptying the Water Trap Chamber
The water trap chamber should be emptied after every test, or when the water
condensate approaches the tip of the riser tube.
1. Remove water trap chamber (also see page 49).
2. Pour out liquid condensate and then reassemble trap.
7.4. Replacing the Filter Element
Replace the filter element when it becomes visibly dirty or becomes saturated
with water.
7.4.1. Material Required
• Filter Element (P/N 0007-1644)
Small Flat Blade Screwdriver
•
7.4.2. Procedure
1. Remove water trap chamber (also see page 49).
2. Pry apart filter chamber using a small flat-blade screwdriver. Remove
and discard old filter.
3. Install new filter and reassemble filter chamber, making sure that
surfaces “A” and “B” contact each other.
4. Reassemble trap.
Figure 7-6. Disassembling the Water Trap/Filter Assembly
0024-9472 Rev 2 83
PCA 3 Manual
7.5. Sensor Replacement
All sensors are replaced in a similar manner. Do the following to replace either
the O , CO , CO , NO, NO , or SO sensor. Refer to Replacement Parts on
2 LOW HIGH 2 2
page 90 for a list of sensors and part numbers.
NOTE: The typical life expectancy of the O , SO and NO sensors
2 2 2
is 2 years. Typical life expectancy of the CO, CO and NO
HIGH
sensors is 3 years.
1. Turn OFF the analyzer.
2. Remove the battery cover.
3. Remove the sensor retainer.
4. Remove the tubing from the gas cup of sensor being replaced.
5. Unplug the sensor along with its printed circuit board from the analyzer’s
main board.
6. Remove the sensor’s gas cup by twisting its bayonet fitting
counterclockwise, and then pulling the cup straight up.
7. Carefully remove the old sensor from the B-Smart® sensor PCB,
keeping in mind that there is an adhesive foam disk in between the PCB
and sensor.
IMPORTANT: When replacing the O sensor, be sure that the “+”
2
sensor pin plugs into the PCB socket that is also marked “O +”.
2
Also, when replacing an NO sensor, replace the bias battery.
8. Remove the replacement sensor from its plastic canister. Note that the
CO , CO , SO , and NO sensors have a shorting spring installed
LOW HIGH 2 2
between two of their pins. Remove and discard this spring before
installing the sensor.
9. Attach the gas cup to new sensor, making sure that when the sensor
assembly is plugged into the analyzer’s main board, the tubing
connections on top of gas cup are aligned as shown in Figure 7-7.
10. Plug the sensor into analyzer, and then reattach tubing to gas cup.
11. Install the sensor retainer and battery cover.
84 0024-9472 Rev 2
PCA 3 Manual
CAUTION:
Figure 7-7. Sensor Assembly
12. Allow the sensor that was just installed time to stabilize in the circuit
before continuing with this procedure. Stabilization time for all sensors
(except for the NO sensor) is about 1 hour. The NO sensor baseline
technically requires several days to stabilize, but should be sufficiently
stabilized for use in approximately 4 hours.
13. Turn ON the analyzer and confirm that no sensor errors occur during
warm up.
NOTE: Discard old sensors in accordance with local and federal
hazardous waste disposal laws.
®
14. Enter the B-Smart calibration code or calibrate the new sensor(s) as
explained earlier (see Calibration on page 62). An exception is the O
2
sensor which does not require calibration.
0024-9472 Rev 2 85
PCA 3 Manual
7.6. Nitric Oxide (NO) Sensor Battery Replacement
A single lithium battery, located on the NO Smart Sensor assembly, applies a
bias voltage to the NO sensor to prevent the sensor from destabilizing when the
analyzer is turned off. The NO bias battery is expected to last at least the life of
the NO sensor.
NOTE: It is recommended that the bias battery be replaced
whenever the NO sensor is replaced.
7.6.1. Material Required
• Bias battery (refer to Replacement Parts on page 90)
7.6.2. Procedure
®
1. Follow the instructions on page 78 (PCA 3 Disassembly) to remove the
sensor assembly from sensor position 4.
2. Remove the old battery from its holder on the nitric oxide printed circuit
board.
3. Insert the new battery with the positive side toward the sensor (the
battery contact is stamped with a + symbol).
4. Re-install the Smart Sensor assembly.
5. Before powering up and using the analyzer, allow the nitric oxide sensor
to stabilize as described below. Note that recalibration of the nitric oxide
sensor is not required after replacing its bias battery.
Depending on how long the nitric oxide sensor was without bias voltage,
the time required for the sensor to completely stabilize varies from less
than a minute to several days. Typical stabilization times are shown
below. Generally, however, the sensor is sufficiently stable after 4 hours
for measurement purposes.
Bias Battery Stabilization Specifications for NO Sensor
Bias Battery Removed for... Stabilization Time
Less than 15 minutes Less than 1 minute
Less than 1 hour Less than 5 minutes
Less than 2 days Less than 4 hours
Greater than 2 days Up to 2 days
86 0024-9472 Rev 2
PCA 3 Manual
7.7. Cleaning the Probe
The probe tube and gas sample hose will become dirty under normal use.
NOTE: The water trap’s filter element should prevent soot from
reaching the analyzer’s internal components. If the probe is not kept
clean, it could become clogged and restrict the flow of gas into the
analyzer, resulting in incorrect combustion test readings and
calculations.
NOTE: An analyzer that is used to test natural gas furnaces
normally requires less frequent cleaning than an analyzer used for
testing coal or oil fired furnaces.
7.7.1. Equipment Required
• Alcohol
• Aerosol Can of Automotive Carburetor Cleaner
• Clean Rag
Source of Compressed Air (optional)
•
7.7.2. Procedure
1. Remove gas sample hose from top of water trap.
CAUTION: Carburetor cleaner attacks plastic
components! Take precautions not to spray cleaner onto
the probe handle or analyzer.
2. Insert the plastic spray tube of the carburetor cleaner into the gas
sample hose, and then liberally spray carburetor cleaner through the
hose and out the probe tube.
3. After spraying, remove all the residual cleaner by repeatedly flushing
the gas hose and probe tube with alcohol.
4. Wipe off the surfaces of the probe and tubing with a clean rag.
5. Allow the parts to dry completely. If available, blow compressed air
through the probe to expedite the drying process.
6. Reconnect gas sample hose to top of water trap.
0024-9472 Rev 2 87
PCA 3 Manual
CHAPTER 8. TROUBLESHOOTING
8.1. Error Symbols
Error symbols are shown in the data fields of the Combustion Test screens and
on the printout to indicate the following.
Error Symbol Description
Three hyphens or dashes indicates that the calculated data cannot
be displayed because the measured data necessary to make the
- - - calculation is out of range. For calculations to occur, the oxygen level
must be below 16% and the stack temperature must be below
2,000 °F (1,093 °C).
Three stars appear in the data field of sensors that are not installed,
defective, or found to be in error during warm-up, and also in the data
fields of that sensor’s related calculated values. For example, if the
* * *
data field of either the NO or NO sensor contains three stars, then
2
the NOx data field will also contain three stars.
X X X Indicates sensor over range.
8.2. Diagnostics and Status Screens
The DIAGNOSTICS menu provides information regarding the operation of the
analyzer. Information includes the following.
Diagnostic Description
Time Meters Displays the run time of the
analyzer, sample pump and
purge pump in hours of
operation.
Figure 8-1. Time Meters
Main Lists the current status of the Stack and Air Thermocouple channels,
Diagnostics Reference Temperature channel, Pressure channel, and Battery.
O Sensor Life Displays the approximate remaining life of the Oxygen sensor.
2
B-Smart Displays the current status of all B-Smart sensors installed.
Sensors
Fresh Air Fresh air diagnostics will cycle the instrument through the 60 second
Diagnostics warm-up to check on the status of all sensors. The probe must be in
fresh air to perform a valid check.
88 0024-9472 Rev 2
PCA 3 Manual
8.3. Accessing the Diagnostic Menu
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
Figure 8-2. Accessing the Diagnostics Menu
2. Use the ▲▼ buttons to highlight DIAGNOSTICS and then press ENT to
display the DIAGNOSTICS MENU.
3. Use the ▲▼ buttons to highlight the desired diagnostic topic, and then
press ENT to display the information under that topic.
4. Press ESC key to exit back to the Diagnostic menu or the Menu (F2)
key to exit back to the Main Menu.
8.4. Accessing the Status Menu
The Status screen provides a quick reference to key items when troubleshooting.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight STATUS and then press ENT to
display the DEVICE STATUS screen.
3. Press ESC key to exit back to the Main Menu.
8.5. Error Messages (After Warm-up)
If there were problems detected during warm-up, error messages that describe
the nature of the problems are displayed immediately following the analyzer’s 60-
second warm-up period.
If problems were detected, the analyzer will not automatically switch to the
Combustion Test HOLD screen after warm-up. The analyzer, however, can still
be used to perform any test that does not depend on the sensor that is in error.
Press the RUN button to manually display the Combustion Test HOLD screen,
and then continue using the analyzer.
The following is a list of the error messages that may appear following warm-up
and their suggested remedies.
0024-9472 Rev 2 89
PCA 3 Manual
Error Message Description
Low Battery Battery voltage is low. Replace batteries (page 20).
O Sensor Oxygen sensor not installed. Install oxygen sensor. Refer to
2
Missing Sensor Replacement on page 84.
BAD Sensors
The oxygen sensor’s output is too low and can not be calibrated
in the instrument, signifying that the sensor is depleted and
needs to be replaced. Refer to Sensor Replacement on page 84.
No B-Smart There are no B-Smart sensor(s) installed. Refer to Sensor
Sensors Installed Replacement on page 84.
Low Sensors O , CO , CO , NO, NO , or SO sensor output(s) were low,
2 LOW HIGH 2 2
but still usable. Sensor(s) may need to be replaced in the near
future. Message will indicate which sensor(s) are in warning.
T-STK The probe’s thermocouple is not connected to the analyzer’s
Disconnected T-STACK connector. Plug the probe thermocouple plug into the
T-STACK connector at the bottom of the instrument.
Warm-up Sensor CO , CO , NO, NO , or SO Gas sensor(s) were not zeroed
LOW HIGH 2 2
Error at warm-up because of high output. Run instrument on fresh air
then restart instrument to re-zero sensor(s). If message persists,
sensor(s) may need to be replaced. Message will indicate which
sensor(s) are in error.
• Stack or Air temperature channel is measuring temperature
outside the range of -4 to 212 °F at startup. Make sure that
the Stack and Air thermocouples are sampling ambient
room air within the temperature range at startup. Message
will indicate which channel(s) are in error.
• Pressure sensor is measuring pressure outside the range of
+/- 3 inches of water column at startup. Ensure that the
analyzer is sampling atmospheric pressure and restart.
The analyzer was turned on with the probe sampling flue
•
gas. Move the probe to fresh air and restart the analyzer.
8.6. Replacement Parts
Item Description Part No.
2 LCD Module 0024-1520
3 Rear Case 0024-1381
4 Top Case 0024-1524
5 Battery Cover 0024-1383
6 Window 0024-1526
7 LCD Frame 0024-1525
8 Keypad 0024-1387
9 Chassis 0024-1388
10 Lens, Infrared 0024-1391
11 Replacement Pump, Sample Gas 0024-1547
90 0024-9472 Rev 2
PCA 3 Manual
Item Description Part No.
12 Replacement Pump, Purge 0024-1548
13 Sensor Retainer 0024-1418
14 Replacement Hose Connector Plate (complete 0024-1480
assembly, including O-rings)
14A O-Ring, Pressure Fitting 0105-5103
14B O-Ring, Gas Fitting 0105-5102
15A Gas Cup, O /SO /NO 0024-1421
2 2 2
15B Gas Cup, CO 0024-1422
LOW
15C Gas Cup, NO/CO 0024-1420
HIGH
16A Sensor, O 0024-0788
2
®
16B Sensor, B-Smart , CO 0024-1541
LOW
®
16C Sensor, B-Smart , CO 0024-1542
HIGH
®
16D Sensor, B-Smart , NO 0024-1545
®
16E Sensor, B-Smart , NO 0024-1544
2
®
16F B-Smart Sensor, SO 0024-1543
2
16G Sensor, NOT calibrated, CO 0024-0789
LOW
16H Sensor, NOT calibrated, CO 0024-0997
HIGH
16J Sensor, NOT calibrated, NO 0024-0881
16K Sensor, NOT calibrated, NO 0024-1027
2
16L Sensor, NOT calibrated, SO 0024-0998
2
17 Screw, #4 x 1/2” LG 0002-2144
18 Battery Clip, Single 0004-1434
19 Battery Clip, Double 0024-1433
20 Cable Assembly 0024-1521
21 Tubing, Vinyl, 1/8 ID x 3/16 OD* 0003-6104
22 Tubing, Silicon, 1/16 ID x 1/8 OD* 0003-6372
23 Tubing, Silicon, 3/32 ID x 5/32 OD* 0103-6101
24 Tubing, Silicon, 1/8 ID x 3/16 OD* 0103-6102
25 Tubing, 1/8 ID x 1/4 OD* 0003-6105
26 Elbow 0003-6170
- - Bias Battery 0204-0020
- - O-Ring Kit (all O-rings included) 0024-1471
0024-9472 Rev 2 91
PCA 3 Manual
Main PCB Assembly
(Replacement Part Not Available)
Figure 8-3. Replacement Parts
92 0024-9472 Rev 2
PCA 3 Manual
Figure 8-3. Replacement Parts (Continued)
8.7. Accessories
Standard Accessories
Description Part No.
Carrying Case 0024-0865
Batteries, ‘AA’ Alkaline 0204-0004
12” Probe, Hose, and Water Trap/Filter Assembly 0024-3004
1 Water Trap/Filter 0019-3265
2 Filter Element (3 pack) 0007-1644
3 Probe Stop 0019-0580
4 Thumb Screw 0102-0875
5 Connector, Gas Sample 0024-0877
6 Connector, Draft 0024-0878
®
Fyrite User Software 0024-1470
USB Cable 0104-4032
Instruction Manual 0024-9472
0024-9472 Rev 2 93
PCA 3 Manual
Figure 8-4. Hose and Probe Assembly (Parts)
Optional Accessories
Description Part No.
AC Power Adapter:
0024-1254
(Input: 100-240 VAC, 50/60 Hz; Output: 6 VDC @ 1 A)
Ambient Air Thermocouple (T-AIR), K-Type, 10 ft. 0104-1797
Ambient Air Thermocouple (T-AIR), K-Type, 1 inch. 0104-1798
Utility Wand (12 in. rigid probe w/ 5 ft. coiled cable) 0104-1799
Differential Pressure Hose Assembly, 6 ft. 0024-1103
Calibration Kit (includes hoses, adapter, flow meter; does not include gas
0024-7059
cylinders)
Gas Cylinder, 500 ppm CO in air (103 Liter) 0024-0492
Gas Cylinder, 1000 ppm CO & 1000 ppm H in Nitrogen (103 Liter) 0024-0794
2
IrDA Printer 0024-1400
Printer Paper, 1 Roll 0006-8733
Printer Paper, 5 Pack 0024-1310
Probe, Hose, and Water Trap/Filter Assembly, 6 in. Probe 0024-3002
Probe, Hose, and Water Trap/Filter Assembly, 24 in. Probe 0024-3021
Probe, Hose, and Water Trap/Filter Assembly, 36 in. Probe 0024-3022
Thermocouple Replacement Kit, 6 in. 0024-8413
Thermocouple Replacement Kit, 12 in. 0024-8414
94 0024-9472 Rev 2
PCA 3 Manual
Description Part No.
Thermocouple Replacement Kit, 24 in. 0024-8415
Thermocouple Replacement Kit, 36 in. 0024-8416
Sample Conditioning Probe, Compact
0024-7224
(recommended when measuring NO and SO )
2 2
8.8. Analyzer Repair
®
It is recommended that field repair of the PCA 3 be limited to:
• Checks of printed circuit board connectors
Replacing the probe assembly
•
• Replacing the filter element in the water trap / filter assembly
Replacing sensors.
•
Information on how to perform these repairs is provided on page 78 (the
Maintenance section). All other repairs should be performed by an authorized
Bacharach Service Center (refer to the Service Centers section on page 95).
Any repairs performed by an unauthorized service organization will void the
analyzer’s warranty and release Bacharach, Inc. of any implied or written product
liability.
8.9. Service Centers
United States
Bacharach, Inc.
621 Hunt Valley Circle
New Kensington, PA 15068
Phone: 724-334-5051
Fax: 724-334-5723
Email: help@MyBacharach.com
Canada
Bacharach of Canada, Inc.
20 Amber Street Unit #7
Markham, Ontario L3R 5P4
Canada
Phone: 905-470-8985
Fax: 905-470-8963
Email: bachcan@idirect.com
0024-9472 Rev 2 95
PCA 3 Manual
CHAPTER 9. DECLARATION OF CONFORMITY
DECLARATION OF CONFORMITY
The manufacturer of the products Bacharach, Inc.
covered by this declaration:
621 Hunt Valley Circle
New Kensington, PA 15068
Year conformity is declared: 2010
Product(s): Combustion Analyzer
Model(s): PCA 3 (Domestic)
The undersigned hereby declares that the above referenced products are in conformity
with the provisions of the following standard(s) and is in accordance with the following
directive(s).
Standard(s):
Electrical Apparatus for the Detection and
Electromagnetic
EN 50270:2006 Measurement of Combustible Gases, Toxic
Compatibility
Gases, or Oxygen
Radio Limits and Methods of Measurements for
EN 55011 Disturbance Industrial, Scientific and Medical (ISM) Radio-
Characteristics Frequency Equipment
Directive(s):
2004/108/EC EMC Directive
Signature:
Name: Doug Keeports
Title: VP of Product Development
Date: 18 October 2010
The technical documentation file required by this directive is maintained at the corporate headquarters of Bacharach, Inc.
96 0024-9472 Rev 2
PCA 3 Manual
0024-9472 Rev 2 97
PCA 3 Manual
98 0024-9472 Rev 2
PCA 3 Manual
0024-9472 Rev 2 99
PCA 3 Manual
World Headquarters
621 Hunt Valley Circle, New Kensington, Pennsylvania 15068
Phone: 724-334-5000 • Toll Free: 1-800-736-4666 • Fax: 724-334-5001
Website: www.MyBacharach.com • E-mail: help@MyBacharach.com
100 0024-9472 Rev 2
What they say about us
FANTASTIC RESOURCE
One of our top priorities is maintaining our business with precision, and we are constantly looking for affiliates that can help us achieve our goal. With the aid of GID Industrial, our obsolete product management has never been more efficient. They have been a great resource to our company, and have quickly become a go-to supplier on our list!
Bucher Emhart Glass
EXCELLENT SERVICE
With our strict fundamentals and high expectations, we were surprised when we came across GID Industrial and their competitive pricing. When we approached them with our issue, they were incredibly confident in being able to provide us with a seamless solution at the best price for us. GID Industrial quickly understood our needs and provided us with excellent service, as well as fully tested product to ensure what we received would be the right fit for our company.
Fuji
HARD TO FIND A BETTER PROVIDER
Our company provides services to aid in the manufacture of technological products, such as semiconductors and flat panel displays, and often searching for distributors of obsolete product we require can waste time and money. Finding GID Industrial proved to be a great asset to our company, with cost effective solutions and superior knowledge on all of their materials, it’d be hard to find a better provider of obsolete or hard to find products.
Applied Materials
CONSISTENTLY DELIVERS QUALITY SOLUTIONS
Over the years, the equipment used in our company becomes discontinued, but they’re still of great use to us and our customers. Once these products are no longer available through the manufacturer, finding a reliable, quick supplier is a necessity, and luckily for us, GID Industrial has provided the most trustworthy, quality solutions to our obsolete component needs.
Nidec Vamco
TERRIFIC RESOURCE
This company has been a terrific help to us (I work for Trican Well Service) in sourcing the Micron Ram Memory we needed for our Siemens computers. Great service! And great pricing! I know when the product is shipping and when it will arrive, all the way through the ordering process.
Trican Well Service
GO TO SOURCE
When I can't find an obsolete part, I first call GID and they'll come up with my parts every time. Great customer service and follow up as well. Scott emails me from time to time to touch base and see if we're having trouble finding something.....which is often with our 25 yr old equipment.
ConAgra Foods